THEORETICAL CONSIDERATIONS ON THE DESIGN OF CYLINDRICAL FLOW CELLS UTILIZING OPTICAL FIBERS

被引:6
作者
BRUNO, AE
KUHN, R
WIDMER, HM
机构
[1] Central Analytical Department, CIBA-GEIGY Ltd., Basle
关键词
D O I
10.1016/S0003-2670(00)83566-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of microbore cylindrical flow cells featuring optical fibres depends on the geometry and materials of the components. The results of an algorithm which optimizes cell transmittance and minimizes inherent refractive index effects are presented. Optimum fibres and capillary tube geometries are expressed in terms of three reduced parameters for general application. © 1990.