ENZYMATIC DETERMINATION OF CITRATE IN PLASMA WITHOUT DEPROTEINISATION
被引:27
作者:
ZENDER, R
机构:
[1] Laboratoire de l'Hôpital de La Chaux-de-Fonds Dir. Dr. R. Zender., Service de Médecine, l'Hôpital de la Chaux-de-Fonds Dir. Dr. B. Courvoisier
关键词:
D O I:
10.1016/0009-8981(69)90102-8
中图分类号:
R446 [实验室诊断];
R-33 [实验医学、医学实验];
学科分类号:
1001 ;
摘要:
A modification of the Moellering method for citrate determination in blood plasma is proposed. The analysis is performed without deproteinisation. It is shown that: 1. (a) the citrate level remains constant in frozen plasma (-20°) during 16 days at least, 2. (b) this level is also constant when plasma is incubated during 8 h at +20°, 3. (c) frequent or normal values for plasma citrate (P = 0.05) are: 9.4 to 25.2 mg/1. The distribution is gaussian. © 1969.