SURFACE DOSE-RATE CALIBRATION OF SR-90 PLANE OPHTHALMIC APPLICATORS

被引:15
作者
GOETSCH, SJ [1 ]
SUNDERLAND, KS [1 ]
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
关键词
CALIBRATION; EXTRAPOLATION CHAMBER; SR-90; BETA-DOSIMETRY; OPHTHALMIC APPLICATORS;
D O I
10.1118/1.596702
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Calibration of an imported strontium-90 ophthalmic applicator at the U.S. National Bureau of Standards (now the National Institute of Standards and Technology) has disclosed a significant discrepancy in dose rate calibration (32% - 35%) with that quoted by the manufacturer. The University of Wisconsin has investigated this discrepancy and found that both laboratories use similar techniques and a version of the Bragg-Gray equation to yield dose rate estimates. Experimental results indicate a strong relationship between the size of the collecting electrode used in the extrapolation chamber and the resulting estimate of absorbed dose rate. Calibration of these applicators is reviewed and suggestions for improvement and further research are proposed.
引用
收藏
页码:161 / 166
页数:6
相关论文
共 14 条
[1]  
AINSWORTH A, 1989, UNPUB WORKSHOP OPHTH
[2]  
BOHM J, 1976, 8 C INT SOC FRANC RA
[3]   BETA-RAY APPLICATION TO THE EYE - WITH THE DESCRIPTION OF AN APPLICATOR UTILIZING SR-90 AND ITS CLINICAL USE [J].
FRIEDELL, HL ;
THOMAS, CI ;
KROHMER, JS .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1950, 33 (04) :525-535
[4]   CALIBRATION OF SR-90 OPHTHALMIC APPLICATORS [J].
GOETSCH, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 16 (06) :1653-1653
[5]   DESIGN AND OPERATION OF AN EXTRAPOLATION CHAMBER WITH REMOVABLE ELECTRODES [J].
LOEVINGE.R ;
TROTT, NG .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1966, 17 (02) :103-&
[6]   CALIBRATION OF BETA-PARTICLE OPHTHALMIC APPLICATORS AT THE NATIONAL-BUREAU-OF-STANDARDS [J].
PRUITT, JS .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1986, 91 (04) :165-170
[7]  
PRUITT JS, 1987, NBS SPECIAL PUBLICAT, V2509
[8]  
PRUITT JS, 1988, COMMUNICATION
[9]   DOSIMETRY OF SR-90 OPHTHALMIC APPLICATORS [J].
REFT, CS ;
KUCHNIR, FT ;
ROSENBERG, I ;
MYRIANTHOPOULOS, LC .
MEDICAL PHYSICS, 1990, 17 (04) :641-646
[10]   THE CONSTRUCTION AND MEASUREMENT OF BETA-RAY APPLICATORS FOR USE IN OPHTHALMOLOGY [J].
SINCLAIR, WK ;
TROTT, NG .