VENA TECH-LGM FILTER - LONG-TERM RESULTS OF A PROSPECTIVE-STUDY

被引:91
作者
CROCHET, DP
STORA, O
FERRY, D
GROSSETETE, R
LEURENT, B
BRUNEL, P
NGUYEN, JM
机构
[1] Department of Thoracic Radiology, Hemodynamic Center, Hop. G. et R. Laennec-CHR Nantes
关键词
VENAE-CAVAE; FILTERS; INTERVENTIONAL PROCEDURE; STENOSIS OR OBSTRUCTION;
D O I
10.1148/radiology.188.3.8351362
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Efficiency and tolerance of the Vena Tech-LGM filter were assessed in 142 patients prospectively studied during a 2-6-year follow-up period. No patients were lost to follow-up, and 137 (91.3%) scheduled control examinations were performed. Fifty-three patients died of causes unrelated to thromboembolic disease. Pulmonary embolism was suspected in five surviving patients (3.5%). Sixteen retractions (18.4%), 16 intracaval migrations (18.4%), and one tilt were observed. Filter patency studied at Doppler ultrasonography, with or without phlebocavography, gave the following results according to the Kaplan-Meier actuarial method: 92% patency after 2 years of follow-up, 80% after 4 years, and 70% after 6 years. Vena Tech-LGM filter obstruction was related to retraction (P < 10(-6)) or distal migration (P < .004). Occurrence of trophic disease in the lower limbs during the follow-up period was related to the initial level of deep venous thrombosis (P = .03) and had no relation to the patency of the filter. The filter was effective in preventing pulmonary embolism, with 70% demonstrating long-term patency, and no deleterious effects were imputed to obstructions observed during follow-up.
引用
收藏
页码:857 / 860
页数:4
相关论文
共 22 条
[1]  
BERLAND LL, 1980, AM J ROENTGENOL, V134, P1047, DOI 10.2214/ajr.134.5.1047
[2]   THE CLINICAL COURSE OF PULMONARY-EMBOLISM [J].
CARSON, JL ;
KELLEY, MA ;
DUFF, A ;
WEG, JG ;
FULKERSON, WJ ;
PALEVSKY, HI ;
SCHWARTZ, JS ;
THOMPSON, BT ;
POPOVICH, J ;
HOBBINS, TE ;
SPERA, MA ;
ALAVI, A ;
TERRIN, ML .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (19) :1240-1245
[3]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS [J].
DORFMAN, GS .
RADIOLOGY, 1990, 174 (03) :987-992
[4]   EXPERIENCE WITH THE AMPLATZ RETRIEVABLE VENA-CAVA FILTER [J].
EPSTEIN, DH ;
DARCY, MD ;
HUNTER, DW ;
COLEMAN, CC ;
TADAVARTHY, SM ;
MURRAY, PD ;
CASTANEDAZUNIGA, WR ;
AMPLATZ, K .
RADIOLOGY, 1989, 172 (01) :105-110
[5]   MIGRATION OF A GREENFIELD FILTER TO THE PULMONARY-ARTERY - A CASE-REPORT [J].
FRIEDELL, ML ;
GOLDENKRANZ, RJ ;
PARSONNET, V ;
ALPERT, J ;
BRIEF, DK ;
BRENER, BJ ;
AUERON, FM .
JOURNAL OF VASCULAR SURGERY, 1986, 3 (06) :929-931
[6]   RECURRENT PULMONARY-EMBOLISM AFTER GREENFIELD FILTER PLACEMENT [J].
GEISINGER, MA ;
ZELCH, MG ;
RISIUS, B .
RADIOLOGY, 1987, 165 (02) :383-384
[7]  
GHIDALIA S, 1990, DIAGN INTERVENT RADI, V2, P215
[8]  
GREENFIELD LJ, 1988, SURGERY, V104, P706
[9]  
GREENFIELD LJ, 1981, ARCH SURG-CHICAGO, V116, P1451
[10]  
GREENFIELD LJ, 1991, J VASC SURG, V14, P253