BLOOD-GLUCOSE MEASUREMENT BY MULTIPLE ATTENUATED TOTAL REFLECTION AND INFRARED-ABSORPTION SPECTROSCOPY

被引:39
作者
MENDELSON, Y
CLERMONT, AC
PEURA, RA
LIN, BC
机构
[1] WORCESTER POLYTECH INST, ELECT ENGN, WORCESTER, MA 01609 USA
[2] UNIV MASSACHUSETTS, SCH MED, BIOMED ENGN, WORCESTER, MA 01605 USA
[3] JOSLIN DIABET CTR, WILLIAM P BEETHAM EYE INST, BOSTON, MA USA
关键词
D O I
10.1109/10.55636
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The difficulty of measuring physiological concentrations of glucose in blood by conventional infrared absorption spectroscopy is due to the intrinsic high background absorption of water. This limitation can be largely overcome by the use of a C02 laser as an infrared source in combination with a multiple attenuated total reflection (ATR) technique. To demonstrate the applicability of this technique, we compared in vitro measurements of glucose in blood obtained from an experimental infrared laser spectrometer with independent measurements made by a standard YSI 23A laboratory glucose analyzer. The capability of continuous measurement of blood glucose concentration is of primary importance in the future development of a glucose sensor for diabetic patients. © 1990 IEEE