MEASUREMENT OF END-EXPIRATORY PRESSURE DURING TRANSTRACHEAL HIGH-FREQUENCY JET VENTILATION FOR LARYNGOSCOPY

被引:26
作者
BOURGAIN, JL [1 ]
DESRUENNES, E [1 ]
COSSET, MF [1 ]
MAMELLE, G [1 ]
BELAICHE, S [1 ]
TRUFFABACHI, J [1 ]
机构
[1] INST GUSTAVE ROUSSY,DEPT CHIRURG CERVICOFACIALE,F-94800 VILLEJUIF,FRANCE
关键词
Monitoring: airway pressure; Ventilation: high frequency jet; transtracheal;
D O I
10.1093/bja/65.6.737
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
An anaesthetic technique using high frequency jet ventilation has been proposed for direct laryngoscopy, but this may expose the patients to the risk of barotrauma. In order to assess this risk, we have measured end-expiratory airway pressure (EEP) through the injector using two three-way solenoid valves mounted in series. At the end of insufflation the first valve was switched off and the apparatus deadspace connected to atmosphere through a large exit port during an adjustable time (decompression time). Then the second valve was switched off and the injection line connected to a transducer, allowing measurement of EEP through the injector. The accuracy of this measurement was tested against airway pressure measured directly in the trachea (Pt) in a lung model. Provided that the decompression time was long enough (70 ms) and the apparatus deadspace was small (6 ml), the difference between EEP and Pt was less than 1 cm H2O for frequencies up to 5 Hz. A clinical evaluation was performed in 64 patients under general anaesthesia before laryngoscopy. EEP correlated with end-expiratory pulmonary volume above apnoeic FRC inferred from abdominal and thoracic displacements. At jet frequencies up to 5 Hz, the correlations between these two variables were satisfactory (r > 0.88), suggesting that EEP is a good indicator of pulmonary overdistension. © 1990 Copyright: 1990 British Journal of Anaesthesia.
引用
收藏
页码:737 / 743
页数:7
相关论文
共 19 条
[1]  
BABINSKI M, 1980, ANESTHESIOLOGY, V52, P178
[2]   RIB CAGE VERSUS ABDOMINAL DISPLACEMENT IN RABBITS DURING FORCED-OSCILLATIONS TO 30 HZ [J].
BOYNTON, BR ;
FREDBERG, JJ ;
BUCKLEY, BG ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 63 (01) :309-314
[3]   PERCUTANEOUS JET VENTILATION [J].
CARDEN, E ;
BECKER, G ;
HAMOOD, H .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1976, 85 (05) :652-655
[4]   TECHNICAL ASPECTS AND CLINICAL IMPLICATIONS OF HIGH-FREQUENCY JET VENTILATION WITH A SOLENOID VALVE [J].
CARLON, GC ;
MIODOWNIK, S ;
RAY, C ;
KAHN, RC .
CRITICAL CARE MEDICINE, 1981, 9 (01) :47-50
[5]   UNILATERAL PNEUMOTHORAX FOLLOWING JET VENTILATION DURING GENERAL-ANESTHESIA [J].
CHANG, JL ;
BLEYAERT, A ;
BEDGER, R .
ANESTHESIOLOGY, 1980, 53 (03) :244-246
[6]   VENTURI JET VENTILATION FOR MICROLARYNGOSCOPY - TECHNIQUE, COMPLICATIONS, PITFALLS [J].
CROCKETT, DM ;
SCAMMAN, FL ;
MCCABE, BF ;
LUSK, RP ;
GRAY, SD .
LARYNGOSCOPE, 1987, 97 (11) :1326-1330
[7]  
DESRUENNES E, 1988, Anesthesiology (Hagerstown), V69, pA250, DOI 10.1097/00000542-198809010-00249
[8]   BAROTRAUMA AND HYPOTENSION RESULTING FROM JET VENTILATION IN CRITICALLY ILL PATIENTS [J].
EGOL, A ;
CULPEPPER, JA ;
SNYDER, JV .
CHEST, 1985, 88 (01) :98-102
[9]   PERIPHERAL AIRWAY PRESSURE DURING HIGH-FREQUENCY VENTILATION [J].
ELIASEN, K ;
MOGENSEN, T ;
ANDERSEN, JB .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1986, 30 (01) :97-100
[10]  
FRANTZ ID, 1985, AM REV RESPIR DIS, V131, P134