APPLICATION OF THE THERMOLUMINESCENT DOSIMETRY TO A PION BEAM

被引:3
作者
VALLEY, JF
GUERID, A
LERCH, P
PACHE, G
LIMA, VD
VINCKENBOSCH, M
机构
[1] Institut de Radiophysique Appliquée, Lausanne
关键词
D O I
10.1007/BF01358282
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The thermoluminescent dosimetry is currently applied to radiation therapy in order to measure dose distributions on phantoms or to control the dose in vivo for special procedures or critical organs. The application of this method to the SIN's (Switzerland) pion beam has been studied. Measurements have been made on the following thermoluminescent materials:7LiF,6LiF, Li(nat)F, Li2B4O7, CaF2 : Mn, and CaF2 : Dy. Except for the last one a reduction (10%) of the sensitivity of the phosphors has been observed in the peak relative to the plateau region. An increase of the6LiF sensitivity relative to that of7LiF has been pointed out due to the neutron background through the (n, α) reaction on6Li. Glow curves of the LiF phosphors have been measured. A reduction (5%) of the intensities of peak 3 and 4 relative to peak 5 has been observed for irradiation in the peak region. This effect however can't be used in order to localize the dosimeter in the beam because the intensity of the thermoluminescence peaks shows a larger variability in a batch of dosimeters. © 1979 Springer-Verlag.
引用
收藏
页码:225 / 229
页数:5
相关论文
共 5 条
[1]  
Becker K., 1973, SOLID STATE DOSIMETR
[2]  
Berger M. J., 1966, SP3036 NASA
[3]  
PATRICK JW, 1976, HEALTH PHYS, V30, P295
[4]   APPROACH TO AN ANALYSIS OF ENERGY RESPONSE OF LIF-TLD TO HIGH-ENERGY ELECTRONS [J].
SHIRAGAI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03) :490-499
[5]  
VALLEY JF, 1976, ATOMKERNENERG/KERNT, V27, P197