A FRAMELESS STEREOTAXIC APPROACH TO NEUROSURGICAL PLANNING BASED ON RETROSPECTIVE PATIENT-IMAGE REGISTRATION

被引:72
作者
TAN, KK
GRZESZCZUK, R
LEVIN, DN
PELIZZARI, CA
CHEN, GTY
ERICKSON, RK
JOHNSON, D
DOHRMANN, GJ
机构
[1] UNIV CHICAGO HOSP & CLIN,DEPT RADIAT ONCOL,CHICAGO,IL 60637
[2] UNIV CHICAGO HOSP & CLIN,DEPT SURG,CHICAGO,IL 60637
关键词
BRAIN NEOPLASM; MAGNETIC RESONANCE IMAGING; STEREOTAXIS; INSTRUMENTATION;
D O I
10.3171/jns.1993.79.2.0296
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
A frameless stereotactic device interfacing an electromagnetic three-dimensional (3-D) digitizer to a computer workstation is described. The patient-image coordinate transformation was found by retrospectively registering a digitizer-derived model of the patient's scalp with a magnetic resonance (MR) imaging-derived model of the same surface. This procedure was performed with routine imaging data, eliminating the need to obtain special-purpose MR images with fiducial markers in place. After patient-image fusion was achieved, a hand-held digitizing stylus was moved over the scalp and tracked in real time on cross-sectional and 3-D brain images on the computer screen. This device was used for presurgical localization of lesions in 10 patients with meningeal and superficial brain tumors. The results suggest that the system is accurate enough (typical error range 3 to 8 mm) to enable the surgeon to reduce the craniotomy to one-half the size advisable with conventional qualitative presurgical planning.
引用
收藏
页码:296 / 303
页数:8
相关论文
共 22 条
[1]  
CRZESZCZUK R, 1990, RADIOLOGY, V177, P290
[2]   3-DIMENSIONAL IMAGING [J].
FISHMAN, EK ;
MAGID, D ;
NEY, DR ;
CHANEY, EL ;
PIZER, SM ;
ROSENMAN, JG ;
LEVIN, DN ;
VANNIER, MW ;
KUHLMAN, JE ;
ROBERTSON, DD .
RADIOLOGY, 1991, 181 (02) :321-337
[3]   RAPID NMR IMAGING OF DYNAMIC PROCESSES USING THE FLASH TECHNIQUE [J].
FRAHM, J ;
HAASE, A ;
MATTHAEI, D .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (02) :321-327
[4]   RETROSPECTIVE FUSION OF RADIOGRAPHIC AND MR DATA FOR LOCALIZATION OF SUBDURAL ELECTRODES [J].
GRZESZCZUK, R ;
TAN, KK ;
LEVIN, DN ;
PELIZZARI, CA ;
HU, XP ;
CHEN, GTY ;
BECK, RN ;
CHEN, CT ;
COOPER, M ;
MILTON, J ;
SPIRE, JP ;
TOWLE, VL ;
DOHRMANN, GJ ;
ERICKSON, RK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (05) :764-773
[5]  
GRZESZCZUK R, 1990, J COMPUT ASSIST TOMO, V177, P290
[6]  
GUTHRIE BL, 1991, NEUROL SURG, V2, P1
[7]   3-DIMENSIONAL MAGNETIC-RESONANCE IMAGES OF THE BRAIN - APPLICATION TO NEUROSURGICAL PLANNING [J].
HU, XP ;
TAN, KK ;
LEVIN, DN ;
GALHOTRA, S ;
MULLAN, JF ;
HEKMATPANAH, J ;
SPIRE, JP .
JOURNAL OF NEUROSURGERY, 1990, 72 (03) :433-440
[8]   A FRAMELESS, ARMLESS NAVIGATIONAL SYSTEM FOR COMPUTER-ASSISTED NEUROSURGERY [J].
KATO, A ;
YOSHIMINE, T ;
HAYAKAWA, T ;
TOMITA, Y ;
IKEDA, T ;
MITOMO, M ;
HARADA, K ;
MOGAMI, H .
JOURNAL OF NEUROSURGERY, 1991, 74 (05) :845-849
[9]  
Kelly PJ, 1991, TUMOR STEREOTAXIS
[10]   SURFACE OF THE BRAIN - 3-DIMENSIONAL MR IMAGES CREATED WITH VOLUME RENDERING [J].
LEVIN, DN ;
HU, XP ;
TAN, KK ;
GALHOTRA, S .
RADIOLOGY, 1989, 171 (01) :277-280