COULOMETRIC FLOW ANALYZER FOR USE WITH IMMOBILIZED ENZYME REACTORS

被引:49
作者
ADAMS, RE [1 ]
CARR, PW [1 ]
机构
[1] UNIV GEORGIA, DEPT CHEM, ATHENS, GA 30602 USA
关键词
D O I
10.1021/ac50029a030
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow coulometric immobilized enzyme analyzer based on the use of a fast totally electrochemical pH-stat was developed. The insolubilized urease had a Km of 43 mM and an activity of 1000-1200 units/g of glass at its pH optimum of 6.8 in 0.2 M sodium perchlorate. In the absence of phosphate buffer, urea adsorbed on the immobilized enzyme, causing broad peaks. Peak width was decreased by adding 1,3-diaminopropane to the electrolyte. It serves as a competitive inhibitor (Ki = 79 .mu.M) and blocks adsorption of urea. Urea could be analyzed with good precision (3%) in simulated sera and in quality control reference sera. Recoveries of urea added to quality control sera were 100%. The major problem involved in determining urea in human serum was a high and irreproducible blank which limited the precision to about 5%.
引用
收藏
页码:944 / 950
页数:7
相关论文
共 45 条
  • [1] ACORD W, 1974, CLIN CHEM, V20, P903
  • [2] ADAMS RN, 1976, ANAL CHEM, V48, P1126, DOI 10.1021/ac50007a045
  • [3] QUANTITATIVE-DETERMINATION OF BLOOD-GLUCOSE USING ENZYME INDUCED CHEMILUMINESCENCE OF LUMINOL
    BOSTICK, DT
    HERCULES, DM
    [J]. ANALYTICAL CHEMISTRY, 1975, 47 (03) : 447 - 452
  • [4] Bowers L D, 1976, Anal Chem, V48, p544A
  • [5] BOWERS LD, 1976, CLIN CHEM, V22, P1427
  • [6] NOVEL IMPEDANCE MEASUREMENTS ON ION-SELECTIVE LIQUID-MEMBRANE ELECTRODES
    BRAND, MJD
    RECHNITZ, GA
    [J]. ANALYTICAL CHEMISTRY, 1969, 41 (10) : 1185 - &
  • [7] PREPARATION OF SEVERAL NEW NYLON TUBE GLUCOSE OXIDASE DERIVATIVES AND THEIR INCORPORATION INTO REAGENTLESS AUTOMATED ANALYSIS OF GLUCOSE
    CAMPBELL, J
    HORNBY, WE
    MORRIS, DL
    [J]. BIOCHIMICA ET BIOPHYSICA ACTA, 1975, 384 (02) : 307 - 316
  • [8] CANNING LM, 1975, ANAL LETT, V8, P359
  • [9] CHANEY AL, 1962, CLIN CHEM, V8, P130
  • [10] IMPROVED PENICILLIN SELECTIVE ENZYME ELECTRODE
    CULLEN, LF
    RUSLING, JF
    SCHLEIFER, A
    PAPARIELLO, GJ
    [J]. ANALYTICAL CHEMISTRY, 1974, 46 (13) : 1955 - 1961