EVALUATION OF A SYMMETRICALLY DISPOSED PITOT-TUBE FLOWMETER FOR MEASURING GAS-FLOW DURING EXERCISE

被引:50
作者
PORSZASZ, J [1 ]
BARSTOW, TJ [1 ]
WASSERMAN, K [1 ]
机构
[1] UNIV CALIF LOS ANGELES,LOS ANGELES CTY HARBOR MED CTR,ST JOHNS CARDIOVASC RES CTR,DEPT MED,TORRANCE,CA 90502
关键词
CARDIOPULMONARY EXERCISE TESTING; BREATH-BY-BREATH GAS EXCHANGE;
D O I
10.1152/jappl.1994.77.6.2659
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
We evaluated the effect of airflow and gas composition on the linearity of measurement of airflow by a new disposable flowmeter. The flowmeter is based on the principle of differential pressure measurement across two symmetrically disposed Pitot tubes. Nonlinearities arising from the pressure-to-airflow relationship and sensitivity to changes in gas density were linearized with appropriate software and monitoring of the gas composition. With room air used as the respired gas, the measured tidal volume from a piston pump assembly was consistently within 1-2% of the target tidal volume for each of five flowmeters tested across physiological ranges of flow. Changing gas densities by varying concentrations of O-2, CO2, and N-2 led to errors in tidal volume measurement that ranged up to 6-8%. However, because the errors were predictable, they were corrected by software to within 0.6% of the target volume. Measurement of minute ventilation during exercise was within 1-2% of that determined from bag collections. We conclude that this type of flowmeter can accurately measure exercise minute ventilation and has advantages over some other flowmeters because of its ruggedness, reproducibility, and ease of sterilization or replacement compared with other flowmeters.
引用
收藏
页码:2659 / 2665
页数:7
相关论文
共 9 条
[1]   ON-LINE COMPUTER ANALYSIS AND BREATH-BY-BREATH GRAPHICAL DISPLAY OF EXERCISE FUNCTION TESTS [J].
BEAVER, WL ;
WASSERMAN, K ;
WHIPP, BJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1973, 34 (01) :128-132
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   KINETICS OF OXYGEN-UPTAKE AND HEART-RATE AT ONSET OF EXERCISE IN CHILDREN [J].
COOPER, DM ;
BERRY, C ;
LAMARRA, N ;
WASSERMAN, K .
JOURNAL OF APPLIED PHYSIOLOGY, 1985, 59 (01) :211-217
[4]  
HUSZCZUK A, 1990, EUR RESPIR J, V3, P465
[5]   WATER-VAPOR PRESSURE IN EXPIRED AIR [J].
LIESE, W ;
WARWICK, WJ ;
CUMMING, G .
RESPIRATION, 1974, 31 (03) :252-261
[6]   VENTILATORY RESPONSE TO THE ONSET OF PASSIVE AND ACTIVE EXERCISE IN HUMAN-SUBJECTS [J].
MIYAMURA, M ;
ISHIDA, K ;
YASUDA, Y .
JAPANESE JOURNAL OF PHYSIOLOGY, 1992, 42 (04) :607-615
[7]  
POTHOFF G, 1991, COMPUTERIZED CARDIOPULMONARY EXERCISE TESTING, P183
[8]   THE USE OF A SIMPLE EXERCISE TEST FOR THE DIAGNOSIS OF PNEUMOCYSTIS-CARINII PNEUMONIA IN PATIENTS WITH AIDS [J].
STOVER, DE ;
GREENO, RA ;
GAGLIARDI, AJ .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (06) :1343-1346
[9]   TURBINE FLOWMETER VS FLEISCH PNEUMOTACHOMETER - A COMPARATIVE-STUDY FOR EXERCISE TESTING [J].
YEH, MP ;
ADAMS, TD ;
GARDNER, RM ;
YANOWITZ, FG .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 63 (03) :1289-1295