INVESTIGATION OF THE RELATIVE SURFACE DOSE FROM LIPOWITZ-METAL TISSUE COMPENSATORS FOR 24-MV AND 6-MV PHOTON BEAMS

被引:11
作者
CARDARELLI, GA [1 ]
RAO, SN [1 ]
CAIL, D [1 ]
机构
[1] UNIV HOSP BOSTON, DEPT RADIAT PHYS, BOSTON, MA 02118 USA
关键词
D O I
10.1118/1.596671
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Our radiation therapy department has acquired a tissue compensator system for construction of patient-specific Lipowitz-metal tissue compensators. Since the arrival of this apparatus its use has been limited due to the observance of marked skin reactions located directly beneath the area of the compensator. Although there has been extensive literature on the dose distribution of these metal compensators, there is little data on the effects to the skin for energies greater than 10 MV. Determination of the relative surface dose from Lipowitz-metal tissue compensators is herein investigated for 24- and 6-MV x-ray beams. Effects of field size, source-to-skin distance and thickness of compensator are evaluated as well as the effect of the Lexan tray which supports the compensator during treatment.
引用
收藏
页码:282 / 287
页数:6
相关论文
共 21 条
[1]   USE OF A STYROFOAM BLOCK CUTTER TO MAKE TISSUE-EQUIVALENT COMPENSATORS [J].
BECK, GG ;
MCGONNAGLE, WJ ;
SULLIVAN, CA .
RADIOLOGY, 1971, 100 (03) :694-+
[2]  
Biggs P J, 1979, Med Phys, V6, P291, DOI 10.1118/1.594580
[3]   TISSUE COMPENSATORS FOR MEGAVOLTAGE RADIOTHERAPY FABRICATED FROM HOLLOWED STYROFOAM FILLED WITH WAX [J].
BOGE, RJ ;
EDLAND, RW ;
MATTHES, DC .
RADIOLOGY, 1974, 111 (01) :193-198
[4]   CALCULATION OF DOSE IN HOMOGENEOUS PHANTOMS FOR PARTIALLY ATTENUATED PHOTON BEAMS [J].
ELKHATIB, E ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1988, 15 (02) :145-150
[5]   BROAD BEAM AND NARROW BEAM ATTENUATION IN LIPOWITZS METAL [J].
ELKHATIB, EE ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1987, 14 (01) :135-139
[6]   THE EFFECT OF LEAD ATTENUATORS ON DOSE IN HOMOGENEOUS PHANTOMS [J].
ELKHATIB, EE ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1986, 13 (06) :928-935
[7]  
EVANS RE, 1955, INT SERIES PURE APPL, P712
[8]  
HENDERSON SD, 1983, 25TH ANN M AM ASS PH
[9]  
JOHNS HE, 1983, PHYSICS RAD, P113
[10]   EFFECT OF VARIOUS ATOMIC NUMBER ABSORBERS ON SKIN DOSE FOR 10-MEV X-RAYS [J].
KHAN, FM ;
MOORE, VC ;
LEVITT, SH .
RADIOLOGY, 1973, 109 (01) :209-212