COOLED VIDEO CAMERA FOR OPTICAL INVESTIGATIONS BELOW 1-MK

被引:20
作者
ALLES, H
RUUTU, JP
BABKIN, AV
HAKONEN, PJ
MANNINEN, AJ
PEKOLA, JP
机构
[1] Low Temperature Laboratory, Helsinki University of Technology
关键词
D O I
10.1063/1.1144827
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An optical imaging system for milliKelvin temperatures has been developed based on a regular B/W surveillance camera (25 frames/s), with its CCD sensor inside the 4-K vacuum can of our nuclear demagnetization cryostat. The heat leak to the nuclear stage, caused by the operation of the video camera, was reduced below 1 nW by careful rf shielding. The construction of the system and its limits of operation are discussed.
引用
收藏
页码:1784 / 1785
页数:2
相关论文
共 6 条
[1]  
ALLES H, IN PRESS JLTP
[2]   A PRESTUDY OF OPTICAL IMAGING ON SUPERFLUID HE-3 [J].
MANNINEN, AJ ;
ALLES, H ;
BABKIN, AV ;
PEKOLA, JP .
PHYSICA B, 1992, 178 (1-4) :352-353
[3]   1ST OPTICAL OBSERVATIONS OF SUPERFLUID HE-3 [J].
MANNINEN, AJ ;
PEKOLA, JP ;
KIRA, GM ;
RUUTU, JP ;
BABKIN, AV ;
ALLES, H ;
LOUNASMAA, OV .
PHYSICAL REVIEW LETTERS, 1992, 69 (16) :2392-2395
[4]   SPECTRAL DIFFUSION OF OPTICAL-TRANSITIONS IN DOPED POLYMER GLASSES BELOW 1K [J].
MULLER, KP ;
HAARER, D .
PHYSICAL REVIEW LETTERS, 1991, 66 (18) :2344-2347
[5]   STEPPED SURFACES OF HCP HE-4 CRYSTALS [J].
ROLLEY, E ;
CHEVALIER, E ;
GUTHMANN, C ;
BALIBAR, S .
PHYSICAL REVIEW LETTERS, 1994, 72 (06) :872-875
[6]  
WAGNER R, IN PRESS JLTP