TORQUE FREE ASYMMETRIC GRADIENT COILS FOR ECHO-PLANAR IMAGING

被引:41
作者
ABDULJALIL, AM [1 ]
ALETRAS, AH [1 ]
ROBITAILLE, PML [1 ]
机构
[1] OHIO STATE UNIV,DEPT RADIOL,COLUMBUS,OH 43210
关键词
GRADIENT COIL; ECHO PLANAR IMAGING; MAGNETIC RESONANCE IMAGING;
D O I
10.1002/mrm.1910310415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work, we present a new torque-free asymmetric gradient coil that is capable of generating high quality axial, sagittal, and coronal echo planar images of the human head. This gradient set was calculated using the Biot-Savart law and least square approaches to optimize the field in the region of interest and to minimize net torques. The resulting structure has excellent shoulder-to-coil clearance and, as such, has the advantage of providing good patient access to the linear portion of the gradient while minimizing patient discomfort from claustrophobia.
引用
收藏
页码:450 / 453
页数:4
相关论文
共 13 条
  • [1] ALSOP DC, 1993, 12TH P ANN M NEW YOR, P1359
  • [2] [Anonymous], 1975, CLASSICAL ELECTRODYN
  • [3] [Anonymous], 1986, NUMERICAL RECIPES
  • [4] MAGNETIC-FIELD GRADIENT COILS FOR NMR IMAGING
    BANGERT, V
    MANSFIELD, P
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1982, 15 (02): : 235 - 239
  • [5] HAYES CE, 1986, 5TH P SMRM ANN M, P39
  • [6] PETROPOULOS LS, 1993, 12TH P SMRM ANN M, P1305
  • [7] Quine R. W., 1992, Computers in Physics, V6, P656, DOI 10.1063/1.168422
  • [8] MAGNET FIELD PROFILING - ANALYSIS AND CORRECTING COIL DESIGN
    ROMEO, F
    HOULT, DI
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1984, 1 (01) : 44 - 65
  • [9] DESIGN AND OPTIMIZATION OF A BREAST COIL FOR MAGNETIC-RESONANCE-IMAGING
    SUN, L
    OLSEN, JO
    ROBITAILLE, PML
    [J]. MAGNETIC RESONANCE IMAGING, 1993, 11 (01) : 73 - 80
  • [10] A TARGET FIELD APPROACH TO OPTIMAL COIL DESIGN
    TURNER, R
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1986, 19 (08) : L147 - L151