DESIGN OF A HIGH-PRECISION FRACTION COLLECTOR FOR CAPILLARY ELECTROPHORESIS

被引:53
作者
MULLER, O [1 ]
FORET, F [1 ]
KARGER, BL [1 ]
机构
[1] NORTHEASTERN UNIV,BARNETT INST,BOSTON,MA 02115
关键词
D O I
10.1021/ac00113a036
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
high-precision fraction collector for capillary electrophoresis has been developed. The device utilizes detection close to the end of the capillary and a sheath liquid at the exit of the capillary, allowing continuous collection (i.e., uninterrupted applied electrical field) of multiple species, The role of the sheath liquid now rate and position of detection in the column on the collection precision was assessed. Fiber-optic detection at similar to 1 cm before the exit end of the capillary was found effective for precise timing of the collection. Up to 60 fractions of microliter or smaller volumes could be automatically collected into capillaries used as collection vials. The collection capillaries were placed on a cylinder, and a computer-controlled stepping motor aligned the appropriate capillary with the column exit. The effectiveness of the fraction collector was demonstrated in the collection of all 11 fragments of the HaeIII restriction digest of Phi X-174 plasmid DNA. Polymerase chain reaction amplification of the 271 and 281 bp fragments revealed an inversion of the size-dependent migration order.
引用
收藏
页码:2974 / 2980
页数:7
相关论文
共 34 条
  • [1] THE USE OF CAPILLARY ELECTROPHORESIS IN A MICROPREPARATIVE MODE - METHODS AND APPLICATIONS
    ALBIN, M
    CHEN, SM
    LOUIE, A
    PAIRAUD, C
    COLBURN, J
    WIKTOROWICZ, J
    [J]. ANALYTICAL BIOCHEMISTRY, 1992, 206 (02) : 382 - 388
  • [2] PEAK HOMOGENEITY DETERMINATION AND MICROPREPARATIVE FRACTION COLLECTION BY CAPILLARY ELECTROPHORESIS FOR PHARMACEUTICAL ANALYSIS
    ALTRIA, KD
    DAVE, YK
    [J]. JOURNAL OF CHROMATOGRAPHY, 1993, 633 (1-2): : 221 - 225
  • [3] ARLINGER L, 1979, ELECTROPHORESIS SURV
  • [4] DETECTION OF ENZYME-ACTIVITY IN FRACTIONS COLLECTED FROM FREE SOLUTION CAPILLARY ELECTROPHORESIS OF COMPLEX SAMPLES
    BANKE, N
    HANSEN, K
    DIERS, I
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2): : 325 - 335
  • [5] DIRECT ANALYSIS OF PEPTIDES AND AMINO-ACIDS FROM CAPILLARY ELECTROPHORESIS
    BERGMAN, T
    AGERBERTH, B
    JORNVALL, H
    [J]. FEBS LETTERS, 1991, 283 (01) : 100 - 103
  • [6] SEQUENCE-DEPENDENT MIGRATION BEHAVIOR OF DOUBLE-STRANDED DNA IN CAPILLARY ELECTROPHORESIS
    BERKA, J
    PARIAT, YF
    MULLER, O
    HEBENBROCK, K
    HEIGER, DN
    FORET, F
    KARGER, BL
    [J]. ELECTROPHORESIS, 1995, 16 (03) : 377 - 388
  • [7] SEPARATION BY CAPILLARY ELECTROPHORESIS FOLLOWED BY DYNAMIC ELUTION
    CAMILLERI, P
    OKAFO, GN
    SOUTHAN, C
    [J]. ANALYTICAL BIOCHEMISTRY, 1991, 196 (01) : 178 - 182
  • [8] ANALYTICAL AND MICROPREPARATIVE CAPILLARY ELECTROPHORESIS OF THE PEPTIDES FROM CALCITONIN
    CAMILLERI, P
    OKAFO, GN
    SOUTHAN, C
    BROWN, R
    [J]. ANALYTICAL BIOCHEMISTRY, 1991, 198 (01) : 36 - 42
  • [9] LOW-COST, HIGH-SENSITIVITY LASER-INDUCED FLUORESCENCE DETECTION FOR DNA SEQUENCING BY CAPILLARY GEL-ELECTROPHORESIS
    CHEN, DY
    SWERDLOW, HP
    HARKE, HR
    ZHANG, JZ
    DOVICHI, NJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2): : 237 - 246
  • [10] ISOELECTRIC-FOCUSING BY FREE SOLUTION CAPILLARY ELECTROPHORESIS
    CHEN, SM
    WIKTOROWICZ, JE
    [J]. ANALYTICAL BIOCHEMISTRY, 1992, 206 (01) : 84 - 90