Influence of conductance changes on patch clamp capacitance measurements using a lock-in amplifier and limitations of the phase tracking technique

被引:21
作者
Debus, K [1 ]
Hartmann, J [1 ]
Kilic, G [1 ]
Lindau, M [1 ]
机构
[1] MAX PLANCK INST MED RES,ABT MOLEK ZELLFORSCH,D-69120 HEIDELBERG,GERMANY
关键词
D O I
10.1016/S0006-3495(95)80154-2
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We characterized the influence of conductance changes on whole-cell patch clamp capacitance measurements with a lock-in amplifier and the limitations of the phase-tracking method by numerical computer simulations, error formulas, and experimental tests. At correct phase setting, the artifacts in the capacitance measurement due to activation of linear conductances are small. The cross talk into the capacitance trace is well approximated by the second-order term in the Taylor expansion of the admittance. In the case of nonlinear current-voltage relationships, the measured conductance corresponds to the slope conductance in the range of the sine wave amplitude, and the cross talk into the capacitance trace corresponds to the second-order effect of the slope conductance. The finite gating kinetics of voltage-dependent channels generate phase-shifted currents. These lead to major artifacts in the capacitance measurements when the angular frequency of the sine wave is dose to the kinetic rate constant of the channel. However, when the channel kinetics are sufficiently slow, or sufficiently fast, the cross talk is still close to the second-order effect of the measured conductance. The effects of activation of voltage-dependent currents on the capacitance measurements may be estimated, provided a detailed characterization of the kinetics and voltage dependence is available. A phase error of the lock-in amplifier of a few degrees leads to significant projections. The phase-tracking method can be used to keep the phase aligned only during periods of low membrane conductance. However, nonideal properties of the equivalent circuit, in particular the fast capacitance between the pipette and the bath solutions, may lead to large phase errors when the phase-tracking method is used, depending on the electrical properties of the cell. in this article we provide practical values, setting the range where possible artifacts are below defined limits. For proper evaluation of capacitance measurements, the capacitance and conductance traces should always be displayed together.
引用
收藏
页码:2808 / 2822
页数:15
相关论文
共 35 条
[1]   CALCIUM REQUIREMENTS FOR SECRETION IN BOVINE CHROMAFFIN CELLS [J].
AUGUSTINE, GJ ;
NEHER, E .
JOURNAL OF PHYSIOLOGY-LONDON, 1992, 450 :247-271
[2]   3 POTASSIUM CHANNELS IN RAT POSTERIOR PITUITARY NERVE-TERMINALS [J].
BIELEFELDT, K ;
ROTTER, JL ;
JACKSON, MB .
JOURNAL OF PHYSIOLOGY-LONDON, 1992, 458 :41-67
[3]   CURRENTS THROUGH THE FUSION PORE THAT FORMS DURING EXOCYTOSIS OF A SECRETORY VESICLE [J].
BRECKENRIDGE, LJ ;
ALMERS, W .
NATURE, 1987, 328 (6133) :814-817
[4]   A LOW-COST METHOD FOR RAPID TRANSFER-FUNCTION MEASUREMENTS WITH DIRECT APPLICATION TO BIOLOGICAL IMPEDANCE ANALYSIS [J].
CLAUSEN, C ;
FERNANDEZ, JM .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1981, 390 (03) :290-295
[5]  
Cole K, 1972, MEMBRANES IONS IMPUL
[6]  
DONELLY DF, 1994, BIOPHYS J, V66, P873
[7]   SODIUM AND CALCIUM CHANNELS IN BOVINE CHROMAFFIN CELLS [J].
FENWICK, EM ;
MARTY, A ;
NEHER, E .
JOURNAL OF PHYSIOLOGY-LONDON, 1982, 331 (OCT) :599-635
[8]   PHASE TRACKING - AN IMPROVED PHASE DETECTION TECHNIQUE FOR CELL-MEMBRANE CAPACITANCE MEASUREMENTS [J].
FIDLER, N ;
FERNANDEZ, JM .
BIOPHYSICAL JOURNAL, 1989, 56 (06) :1153-1162
[9]  
Gillis Kevin D., 1995, P155
[10]   A NOVEL CA2+-DEPENDENT STEP IN EXOCYTOSIS SUBSEQUENT TO VESICLE FUSION [J].
HARTMANN, J ;
LINDAU, M .
FEBS LETTERS, 1995, 363 (03) :217-220