IMPROVED AND SIMPLIFIED TISSUE EXTRACTION METHOD FOR QUANTITATING LONG-CHAIN ACYL-COENZYME A THIOESTERS WITH PICOMOLAR DETECTION USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

被引:28
作者
MANGINO, MJ
ZOGRAFAKIS, J
MURPHY, MK
ANDERSON, CB
机构
[1] Department of Surgery, Washington University School of Medicine, St. Louis, MO 63110
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1992年 / 577卷 / 01期
关键词
D O I
10.1016/0378-4347(92)80612-T
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method has been developed that permits rapid and easy tissue extraction of long-chain acyl-coenzyme A (acyl-CoA) thioesters with sensitive quantitation by reversed-phase high-performance liquid chromatography (RP-HPLC). Tissue homogenants are extracted using a reserve Bligh-Dyer technique, and long-chain acyl-CoA esters are harvested in the methanolic aqueous phase. Complex lipids and phospholipids are removed in the chloroform-rich organic Bligh-Dyer second phase, and long-chain acyl-CoA compounds are further purified from the methanolic aqueous Bligh-Dyer first phase on C18 extraction columns after removal of the methanol. The eluted and purified acyl-CoA esters are then quantitated by RP-HPLC using heptadecanoyl-CoA as an internal standard resulting in a detector sensitivity of about 12 pmol. Ten long-chain acyl-CoA esters from C-12:0 to C20:4 were identified and separated from canine renal cortex and murine liver samples. The predominant acyl-CoA peaks from both kidney and liver were 14:0, 16:1, 16:0, 18:1, 18:2 and 20:4. Murine liver also produced 18:0 and all peaks disappeared after alkaline hydrolysis of the samples. This extraction and quantitation technique can successfully be used for tissue samples as small as 20 mg, and many samples can be processed in a short period of time. The simplicity of the extraction procedure and the sensitivity of the assay make this an attractive alternative approach to quantitating long-chain acyl-CoA thioesters from complex biological samples such as tissues.
引用
收藏
页码:157 / 162
页数:6
相关论文
共 11 条
[1]  
CORKEY BE, 1988, METHOD ENZYMOL, V166, P55
[2]   ASSAY OF SHORT-CHAIN ACYL COENZYME-A INTERMEDIATES IN TISSUE-EXTRACTS BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
CORKEY, BE ;
BRANDT, M ;
WILLIAMS, RJ ;
WILLIAMSON, JR .
ANALYTICAL BIOCHEMISTRY, 1981, 118 (01) :30-41
[3]   AMPHIPATHIC METABOLITES AND MEMBRANE DYSFUNCTION IN ISCHEMIC MYOCARDIUM [J].
CORR, PB ;
GROSS, RW ;
SOBEL, BE .
CIRCULATION RESEARCH, 1984, 55 (02) :135-154
[4]   FAT-METABOLISM IN HIGHER-PLANTS - DETERMINATION OF ACYL-ACYL CARRIER PROTEIN AND ACYL COENZYME A IN A COMPLEX LIPID MIXTURE [J].
MANCHA, M ;
STOKES, GB ;
STUMPF, PK .
ANALYTICAL BIOCHEMISTRY, 1975, 68 (02) :600-608
[5]   DETERMINATION OF INDIVIDUAL LONG-CHAIN FATTY ACYL-COA ESTERS IN HEART AND SKELETAL-MUSCLE [J].
MOLAPARASTSALESS, F ;
SHRAGO, E ;
SPENNETTA, TL ;
DONATELLO, S ;
KNEELAND, LM ;
NELLIS, SH ;
LIEDTKE, AJ .
LIPIDS, 1988, 23 (05) :490-492
[6]   RELATIONSHIP BETWEEN CARBOHYDRATE AND LIPID-METABOLISM AND ENERGY-BALANCE OF HEART-MUSCLE [J].
NEELY, JR ;
MORGAN, HE .
ANNUAL REVIEW OF PHYSIOLOGY, 1974, 36 :413-459
[7]   QUANTITATIVE-DETERMINATION OF ACYL CHAIN COMPOSITION OF SUBNANOMOLE AMOUNTS OF CELLULAR LONG-CHAIN ACYL-COENZYME-A ESTERS [J].
PRASAD, MR ;
SAUTER, J ;
LANDS, WEM .
ANALYTICAL BIOCHEMISTRY, 1987, 162 (01) :202-212
[8]   MEASUREMENT OF THE ACYL-COA INTERMEDIATES OF BETA-OXIDATION BY HPLC WITH ONLINE RADIOCHEMICAL AND PHOTODIODE-ARRAY DETECTION - APPLICATION TO THE STUDY OF [U-C-14]HEXADECANOATE OXIDATION BY INTACT RAT-LIVER MITOCHONDRIA [J].
WATMOUGH, NJ ;
TURNBULL, DM ;
SHERRATT, HSA ;
BARTLETT, K .
BIOCHEMICAL JOURNAL, 1989, 262 (01) :261-269
[9]   MITOCHONDRIAL-FUNCTION IN THE HEART [J].
WILLIAMSON, JR .
ANNUAL REVIEW OF PHYSIOLOGY, 1979, 41 :485-506
[10]   EXTRACTION OF TISSUE LONG-CHAIN ACYL-COA ESTERS AND MEASUREMENT BY REVERSE-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
WOLDEGIORGIS, G ;
SPENNETTA, T ;
CORKEY, BE ;
WILLIAMSON, JR ;
SHRAGO, E .
ANALYTICAL BIOCHEMISTRY, 1985, 150 (01) :8-12