DEPTH DISCRIMINATION IN LASER DOPPLER SKIN BLOOD-FLOW MEASUREMENT USING DIFFERENT LASERS

被引:58
作者
OBEID, AN [1 ]
BOGGETT, DM [1 ]
BARNETT, NJ [1 ]
DOUGHERTY, G [1 ]
ROLFE, P [1 ]
机构
[1] UNIV OXFORD,JOHN RADCLIFFE HOSP,BIOENGN UNIT,OXFORD OX3 9DU,ENGLAND
关键词
BIOLOGICAL MATERIALS - Blood - COMPUTERS; MICROCOMPUTER - Medical Applications - LASERS; GAS - Medical Applications - MATHEMATICAL TRANSFORMATIONS - Fast Fourier Transforms - SPECTRUM ANALYSIS;
D O I
10.1007/BF02442302
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The principle of Doppler broadening of the optical spectrum, when laser light is scattered from moving red blood cells (RBC) in skin tissue, has been successfully employed in a number of laser Doppler skin blood flow meters. The technique is particularly suitable for noninvasive, continuous monitoring. There are many instances in routine dermatology where the superficial dermal plexus alone is most affected by the presence of disease (e. g. psoriasis and eczema). To discriminate between total and superficial skin blood flow it is necessary to select a suitable wavelength of laser light which will be preferentially absorbed and/or scattered in the upper dermis. The purpose of this study was to compare the Doppler shift frequency spectra obtained for three different laser wavelengths under the same conditions of flow through a simple physical model. 7 Refs.