DEVELOPMENT OF AN ANTIBODY-BASED AMPEROMETRIC BIOSENSOR TO STUDY THE REACTION OF 7-HYDROXYCOUMARIN WITH ITS SPECIFIC ANTIBODY

被引:16
作者
DEMPSEY, E
OSULLIVAN, C
SMYTH, MR
EGAN, D
OKENNEDY, R
WANG, J
机构
[1] DUBLIN CITY UNIV,SCH CHEM,DUBLIN 9,IRELAND
[2] DUBLIN CITY UNIV,SCH BIOL SCI,DUBLIN 9,IRELAND
[3] NEW MEXICO STATE UNIV,DEPT CHEM,LAS CRUCES,NM 88033
关键词
7-HYDROXYCOUMARIN; DIFFERENTIAL-PULSE VOLTAMMETRY; ANTIBODY-ANTIGEN REACTIONS; IMMUNOSENSOR;
D O I
10.1039/an9931800411
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric biosensor, based on an anti-7-hydroxycoumarin (7-OH-coumarin) antibody immobilized at the surface of a glassy carbon electrode, and contained behind a cellulose dialysis membrane, is described. The electrochemical behaviour of this metabolite at the bare glassy carbon electrode was found to be well defined by using differential-pulse voltammetry, and a 90% decrease in peak height was observed on binding of the antibody to the antigen, which occurs at the electroactive site of 7-OH-coumarin. The kinetics of the antibody-antigen reaction was investigated and a first-order reaction was observed with k = 0.0329 mA min-1. This system provides a novel method for studying antibody specificity and the kinetics of such antibody-antigen interactions.