THE EFFECTS OF PRESSURE AND FLOW ON HEMOLYSIS CAUSED BY BIO-MEDICUS CENTRIFUGAL PUMPS AND ROLLER-PUMPS - GUIDELINES FOR CHOOSING A BLOOD PUMP

被引:43
作者
TAMARI, Y [1 ]
LEESENSIBA, K [1 ]
LEONARD, EF [1 ]
PARNELL, V [1 ]
TORTOLANI, AJ [1 ]
机构
[1] COLUMBIA UNIV,DEPT CHEM ENGN,NEW YORK,NY 10027
关键词
D O I
10.1016/S0022-5223(19)33970-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Two Bio-Medicus BP-50 centrifugal pumps and two roller pumps were tested simultaneously with porcine blood at 21-degrees +/- 1-degrees-C in four in vitro circuits to determine the effect of four combinations of flow and pressure conditions on blood damage. Flows of 300 ml/min (1/4-inch inner-diameter tubing in the roller pump) and 1775 ml/min (1/2-inch inner-diameter tubing in the roller pump) and pressure differences across the pump (DELTAP = outlet pressure - inlet pressure) of 215 mm Hg (n = 6) and 345 mm Hg (n = 5) were examined. The index of hemolysis (milligrams plasma hemoglobin per 100 L blood pumped) for the BP-50 pump was higher at a flow of 300 ml/min than at a flow of 1775 ml/min (p < 0.0002). At 300 ml/min, the index of hemolysis for the BP-50 pump tended to be higher at 345 mm Hg than at 215 mm Hg (mean +/- standard error of the mean, 135 +/- 22 versus 88 +/- 9, p = 0.059). At 1775 ml/min, there was no difference in the index of hemolysis for the BP-50 pump between 215 and 345 mm Hg (37 +/- 7 versus 29 +/- 5, p = 0.32). With the roller pump, the index of hemolysis was higher at a flow of 300 ml/min than at a flow of 1775 ml/min (p < 0.036), but there was no difference in the indexes of hemolysis between 215 and 345 mm Hg at 300 ml/min (60 +/- 9 versus 61 +/- 11, p = 0.93) or at 1775 ml/min (40 +/- 6 versus 36 +/- 6, p = 0.61). Comparison between the two types of pumps showed that the index of hemolysis was significantly higher for the BP-50 than for the roller pump at a flow of 300 ml/min and a DELTAP of 215 mm Hg (88 +/- 9 versus 60 +/- 9, p = 0.009), as well as at a flow of 300 ml/min and a DELTAP of 345 mm Hg (135 +/- 22 versus 61 +/- 11, p = 0.001). At a flow of 1775 ml/min, there was no difference in the index of hemolysis between the two pumps at either pressure condition. Analysis of these data, combined with data reported in other studies in the literature, indicates that the Bio-Medicus pump is expected to yield a lower index of hemolysis than a roller pump when the ratio of pressure difference to flow is less than 0.27 mm Hg/(ml/min) (ie., low pressure, high flow); the opposite is expected when the ratio of pressure difference to flow is greater than 0.27 mm Hg/(ml/min) (i.e, high pressure, low flow).
引用
收藏
页码:997 / 1007
页数:11
相关论文
共 32 条
[1]  
ALLEN JG, 1958, EXTRACORPOREAL CIRCU, P515
[2]  
BELENGER J, 1983, PROG ARTIF ORGANS, P74
[3]  
BERNSTEIN EF, 1967, SURGERY, V61, P432
[4]   FACTORS INFLUENCING ERYTHROCYTE DESTRUCTION IN ARTIFICIAL ORGANS [J].
BERNSTEIN, EF ;
BLACKSHEAR, PL ;
KELLER, KH .
AMERICAN JOURNAL OF SURGERY, 1967, 114 (01) :126-+
[5]  
Bird R B., 2007, TRANSPORT PHENOMENA
[6]  
BLACKSHEAR PL, 1966, T AM SOC ART INT ORG, V12, P113
[7]  
BLACKSHEAR PL, HDB BIOENGINEERING, pCH15
[8]  
DEVORE JL, 1987, PROBABILITY STATISTI, P518
[9]  
ENGLEHARDT H, 1988, P INT WORKSHOP ROTAR
[10]  
Forstrom R., 1969, THESIS U MINNESOTA M