APPROACH OF IMPROVING PRECISION IN ULTRASONIC DOPPLER BLOODSTREAM SPEED MEASUREMENT BY CHAOS-BASED FREQUENCY DETECTING

被引:1
作者
Zhang Shuqing Jin Shijiu LvJiangtao Zhang Liguo Li Jun Electrical Engineering Institute Yanshan University Qinhuangdao China Precision Instrument and Photoelectric Engineering Institute Tianjin University Tianjin China [66004 ,300072 ]
机构
关键词
D O I
暂无
中图分类号
TM935 [频率、波形参数的测量及仪表];
学科分类号
摘要
<正>It is critical for cerebral vascular disease diagnosis through Doppler to detect the maximum and the minimum of the carotid blood flow speed accurately. A kind of Duffing system under an external periodic power with dump is introduced in the letter, numerical analysis is carried out by four-order Runge-Kutta method. An oscillator array is designed according to the frequency of the ultrasonic wave. When the external signals are inputted, computational algorithm is used to scan the array in turn and analyze the result, and the frequency can be determined. Based on the methods above, detecting the carotid blood flow speed accurately is realized. The Signal-to-Noise Ratio (SNR) of-20.23dB is obtained by the result of experiments. In conclusion, the SNR has been improved and the precision of the measured bloodstream speed has been increased, which can be 0.069% to 0.13%.
引用
收藏
页码:457 / 460
页数:4