Fieldable flow injection analysis system with 1,1′-oxalyldiimidazole chemiluminescence detection capable of quantifying acetylcholine

被引:3
作者
Kang, Hui Rak J. [1 ,2 ]
Kang, Kam Chan [1 ,3 ]
Newby, John G. [4 ]
Lee, Ji Hoon [1 ]
机构
[1] Luminescent MD LLC, Hagerstown, MD 21742 USA
[2] Newton N High Sch, Newtonville, MA 02460 USA
[3] Brookline High Sch, Brookline, MA 02445 USA
[4] Washington Cty Hosp, Dept Pathol, Hagerstown, MD 21740 USA
关键词
PEROXYOXALATE CHEMILUMINESCENCE; CAPILLARY-ELECTROPHORESIS; CHOLINE; IMMUNOASSAY; SEPARATION; ENZYMES; MS;
D O I
10.1039/c0ay00440e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We developed a highly sensitive biosensor for quantifying acetylcholine (ACh) using flow injection analysis system with 1,1'-oxalyldiimidazole chemiluminescence (ODI-CL) detection designed based on the principle of liquid core waveguide. ACh in Tris-HCl (pH 8.5) was incubated with the mixture of 1.0 U/ml acetylcholinesterase, 0.5 U/ml choline oxidase, 0.04 U/ml horseradish peroxidase, and 1.0 mu M Amp lex Red in PBS (pH 7.4) for 15 min at room temperature. The concentration of resorufin formed from the consecutive enzyme reactions was proportional to the concentration of ACh in analytical sample. The dynamic range of linear calibration curve (y = 12444x + 11617, R-2 = 0.998) for the quantification of ACh using the biosensor with ODI CL detection was 0.7 similar to 11.3 mu M. The limit of detection (LOD = background noise + 3 sigma) of the biosensor was as low as 0.14 mu M. Based on the results of recovery test and linearity study, finally, we confirmed that FIA system with ODI CL detection is accurate, precise, and reproducible.
引用
收藏
页码:374 / 379
页数:6
相关论文
共 23 条
[1]   Multilayers enzyme-coated carbon nanotubes as biolabel for ultrasensitive chemiluminescence immunoassay of cancer biomarker [J].
Bi, Sai ;
Zhou, Hong ;
Zhang, Shusheng .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (10) :2961-2966
[2]   Development of DNA aptamers for cytochemical detection of acetylcholine [J].
Bruno, John G. ;
Carrillo, Maria P. ;
Phillips, Taylor ;
King, Blythe .
IN VITRO CELLULAR & DEVELOPMENTAL BIOLOGY-ANIMAL, 2008, 44 (3-4) :63-72
[3]   Rapid quantification of melamine in milk using competitive 1,1′-oxalyldiimidazole chemiluminescent enzyme immunoassay [J].
Choi, JooHee ;
Kim, Young-Teck ;
Lee, Ji Hoon .
ANALYST, 2010, 135 (09) :2445-2450
[4]   Acetylcholine in human CSF:: methodological considerations and levels in dementia of Alzheimer type [J].
Frölich, L ;
Dirr, A ;
Götz, ME ;
Gsell, W ;
Reichmann, H ;
Riederer, P ;
Maurer, K .
JOURNAL OF NEURAL TRANSMISSION, 1998, 105 (8-9) :961-973
[5]   Cyanidin-horseradish peroxidase-hydroperoxide reaction system and its application in enzymelinked immunosensing assays [J].
Gong FuChun ;
Li DingZhong ;
Yang Rong ;
Wei JianKe ;
Cao Zhong ;
Tan ShuZhen ;
Tan YaFei .
SCIENCE IN CHINA SERIES B-CHEMISTRY, 2009, 52 (08) :1142-1147
[6]   Near-Infrared Fluorescence Detection of Acetylcholine in Aqueous Solution Using a Complex of Rhodamine 800 and p-Sulfonato-calix[8] arene [J].
Jin, Takashi .
SENSORS, 2010, 10 (03) :2438-2449
[7]   Encapsulation of enzymes within polymer spheres to create optical nanosensors for oxidative stress [J].
Kim, SH ;
Kim, B ;
Yadavalli, VK ;
Pishko, MV .
ANALYTICAL CHEMISTRY, 2005, 77 (21) :6828-6833
[8]   Microfluidic device capable of sensing ultrafast chemiluminescence [J].
Kim, Young-Teck ;
Ko, Seok Oh ;
Lee, Ji Hoon .
TALANTA, 2009, 78 (03) :998-1003
[9]   Capillary Electrophoresis with Electrospray Ionization Mass Spectrometric Detection for Single-Cell Metabolomics [J].
Lapainis, Theodore ;
Rubakhin, Stanislav S. ;
Sweedler, Jonathan V. .
ANALYTICAL CHEMISTRY, 2009, 81 (14) :5858-5864
[10]   Solvent and pH effects on fast and ultrasensitive 1,1′-oxalyldi(4-methyl)imidazole chemiluminescence [J].
Lee, JH ;
Je, J ;
Hur, J ;
Schlautman, MA ;
Carraway, ER .
ANALYST, 2003, 128 (10) :1257-1261