Development of new cleanup method of polychlorinated dibenzo-p-dioxins/dibenzofurans in fish by freezing-lipid filtration

被引:18
作者
Hazardous Substance Research Team, Korea Basic Science Institute, Seoul, 136-701, Korea, Republic of [1 ]
不详 [2 ]
不详 [3 ]
机构
来源
Anal. Chim. Acta | 2006年 / 1卷 / 31-36期
关键词
Filtration - Fisheries - Freezing - Lipids - Liquid chromatography - Solvents;
D O I
10.1016/j.aca.2006.03.010
中图分类号
学科分类号
摘要
Freezing-lipid filtration as a new method has been developed for the rapid determination of polychlorinated dibenzo-p-dioxins/dibenzofurans (PCDDs/Fs) in biological samples. This method can effectively reduce the time of sample pretreatment, labor and amount of solvents compared with conventional methods. By freezing-lipid filtration procedure, about 90% of lipids in extract could be removed without any significant loss of PCDD/Fs. For further cleanup of extracts after freezing-lipid filtration, automatic parallel LC columns including silica gel, alumina and carbon columns were applied. During automatic parallel LC columns cleanup, most of co-extracted interferences such as residue lipids and fatty acids could be eliminated and dioxins could be separated from many other dioxin-like congeners such as polychlorinated biphenyls by this procedure. The extracts after cleanup were analyzed by high-resolution gas chromatography (HRGC)/high-resolution mass spectrometry (HRMS) using an isotope dilution method. The average recoveries and relative standard deviation (R.S.D.) of 17 native congeners in the spiked fish samples at 8-80 pg/g (n = 3) were ranged between 85.3 and 117.2% and 5.7-20.3%, respectively. © 2006 Elsevier B.V. All rights reserved.
引用
收藏
相关论文
empty
未找到相关数据