A new method to measure heart rate with EMFi and PVDF materials

被引:26
作者
Kärki, S. [1 ]
Lekkala, J. [1 ]
机构
[1] Department of Automation Science and Engineering, Tampere University of Technology, Tampere
关键词
Electromechanical film; Heart rate; Polyvinylidenefiuoride;
D O I
10.1080/03091900903067424
中图分类号
R318.08 [生物材料学]; Q [生物科学];
学科分类号
07 ; 0710 ; 0805 ; 080501 ; 080502 ; 09 ;
摘要
In this paper we propose a new simple method to measure the heart rate of a person sitting on a chair or lying in a bed. The heart rate is measured with a thin sensor pad consisting of separate electromechanical film (EMFi) and polyvinylidenefluoride (PVDF) transducers located beneath the leg of chair or bed. This study aims to evaluate the operation of the sensor system with measurements, and also to compare the results provided by the two transducer materials. Based on the results obtained here, the heart rates measured with the transducers mainly corresponded to the values of reference ECG signal. Some minor differences between the heart rate values of PVDF and EMFi appeared, especially in supine position, possible due to the material sensitivities to different force directions. However, to conclude, both materials seem to be convenient for this kind of measurement of heart rate. © 2009 Informa UK Ltd All rights reserved.
引用
收藏
页码:551 / 558
页数:7
相关论文
共 12 条
[1]  
Choi S., Jiang Z., A novel wearable sensor device with conductive fabric and PVDF film for monitoring cardiorespiratory signals, Sensors and Actuators A: Physical, 128, pp. 317-326, (2006)
[2]  
Niizeki K., Nishidate I., Uchida K., Kuwahara M., Unconstrained cardiorespiratory and body movement monitoring system for home care, Medical and Biological Engineering and Computing, 43, pp. 716-724, (2005)
[3]  
Siivola J., New noninvasive piezoelectric transducer for recording of respiration, heart rate and body movements, Medical and Biological Engineering and Computing, 27, pp. 423-424, (1989)
[4]  
Akhbardeh A., Junnila S., Koivuluoma M., Koivistoinen T., Turjanmaa V., Koobi T., Varri A., Towards a heart disease diagnosing system based on force sensitive chair's measurement, biorthogonal wavelets and neural networks, Engineering Applications of Artificial Intelligence, 20, pp. 493-502, (2007)
[5]  
Postolache O., Silva Girao P., Postolache G., Pereira M., Vital signs monitoring system based on EMFi sensors and wavelet analysis, The IEEE Instrumentation and Measurement Technology Conference, (2007)
[6]  
Kim J.-M., Hong J.-H., Cho M.-C., Cha E.-J., Lee T.-S., Wireless biomedical signal monitoring device on wheelchair using noncontact Electro-mechanical Film sensor, The 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, (2007)
[7]  
Rushmer R., Cardiovascular Dynamics, (1976)
[8]  
Technical Manual
[9]  
Paajanen M., Lekkala J., Kirjavainen K., Electro mechanical film (EMFi) - A new multipurpose electret material, Sensors and Actuators A: Physical, 84, pp. 95-102, (2000)
[10]  
Franco S., Design with Operational Amplifiers and Analog Integrated Circuits, (2002)