Histologic studies of failed dental implants: A retrieval analysis of four different oral implant designs

被引:20
作者
Albrektsson, T. [1 ]
Astrand, P. [1 ]
Becker, W. [1 ]
Eriksson, A.R. [1 ]
Lekholm, U. [1 ]
Malmquist, J. [1 ]
Sennerby, L. [1 ]
机构
[1] Biomaterials Group, Department of Handicap Research, University of Gothenburg, Brunnsgatan 2, 413 12 Gothenburg, Sweden
来源
Clinical Materials | 1992年 / 10卷 / 04期
关键词
Alumina - Dental Equipment And Supplies--Dental Prostheses;
D O I
10.1016/0267-6605(92)90015-L
中图分类号
学科分类号
摘要
A retrieval study was performed on four different implant designs. One Tubingen implant showed a thin white line along the interface. One blade-vent implant demonstrated an interfacial soft tissue reaction with inflammation. In analyses of six integral implants it was demonstrated that the hydroxyapatie coat had come loose at certain points along the interface. Four of five retrieved IMZ-implants came from the same patient. These IMZ-implants had been removed in spite of an undistrubed bone anchorage and demonstrated between 5 and 67% of the surface area in direct contact with bone.