ESR/alanine dosimetry of high-energy electrons in radiotherapy

被引:22
作者
Chu, S. [1 ]
Wieser, A. [1 ]
Feist, H. [1 ]
Regulla, D.F. [1 ]
机构
[1] Yonsei Univ, Korea, Republic of
来源
International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes | 1989年 / 40卷 / 10-12期
关键词
Amino Acids - Radiation Effects - Biomedical Engineering - Radiotherapy - Electrons - Magnetic Resonance - Spectroscopy;
D O I
10.1016/0883-2889(89)90030-0
中图分类号
学科分类号
摘要
Dosimetry with detectors of organic material and electron spin resonance (ESR) spectroscopy for readout has been developed metrologically at GSF for high-level photon radiation. The tissue-equivalent probes also appear appropriate for therapy-level dosimetry. Here, results of dose measurements are reported which were performed in a water phantom using high-energy electrons as produced by a linear accelerator. Firstly, the dose range of the ESR/alanine dosimetry system had to be extended down to around 1 Gy. On this basis the response of the alanine dosimeters was determined for electrons at different therapeutic dose levels from less than 0.5 Gy to 10 Gy. Finally depth-dose measurements were carried out for electrons with initial energies up to 21 MeV. Comparisons between ESR/alanine and ion chamber dosimetry were made performing depth-dose measurements in a water phantom to examine the agreement of both methods under field conditions.
引用
收藏
页码:993 / 996
相关论文