Intercomparison of Humidity and Temperature Sensors:GTS1, Vaisala RS80, and CFH

被引:20
作者
卞建春 [1 ]
陈洪滨 [1 ]
Holger VMEL [2 ]
段云俊 [3 ]
宣越健 [1 ]
吕达仁 [1 ]
机构
[1] Key Laboratory of Middle Atmosphere and Global Environment Observation,Institute of Atmospheric Physics, Chinese Academy of Sciences
[2] GRUAN Lead Center, Meteorologisches Observatorium Lindenberg,German Weather Service  3. Kunming National Reference Climatological Station
基金
中国国家自然科学基金;
关键词
GTS1; radiosonde; humidity; dry bias;
D O I
暂无
中图分类号
P412.1 [地面气象观测];
学科分类号
0706 ; 070601 ;
摘要
GTS1 digital radiosonde, developed by the Shanghai Changwang Meteorological Science and Technology Company in 1998, is now widely used in operational radiosonde stations in China. A preliminary comparison of simultaneous humidity measurements by the GTS1 radiosonde, the Vaisala RS80 radiosonde, and the Cryogenic Frostpoint Hygrometer (CFH), launched at Kunming in August 2009, reveals a large dry bias produced by the GTS1 humidity sensor. The average relative dry bias is in the order of 10% below 500 hPa, increasing rapidly to 30% above 500 hPa, and up to 55% at 310 hPa. A much larger dry bias is observed in the daytime, and this daytime effect increases with altitude. The GTS1 radiosonde fails to respond to humidity changes in the upper troposphere, and sometimes even in the middle troposphere. The failure of GTS1 in the middle and upper troposphere will result in significant artificial humidity shifts in radiosonde climate records at stations in China where a transition from mechanical to digital radiosondes has occurred. A comparison of simultaneous temperature observations by the GTS1 radiosonde and the Vaisala RS80 radiosonde suggests that these two radiosondes provide highly reproducible temperature measurements in the troposphere, but produce opposite biases for daytime and nighttime measurements in the stratosphere. In the stratosphere, the GTS1 shows a warm bias (<0.5 K) in the daytime and a relatively large cool bias (-0.2 K to -1.6 K) at nighttime.
引用
收藏
页码:139 / 146
页数:8
相关论文
共 5 条
[1]   L波段电子探空仪相对湿度误差研究及其应用 [J].
姚雯 ;
马颖 ;
徐文静 .
应用气象学报, 2008, (03) :356-361
[2]   杭州L波段和59-701高空探测系统资料对比分析 [J].
朱兰娟 ;
华行祥 .
气象科技, 2007, (05) :750-754
[3]   Pressure and temperature differences between Vaisala RS80 and RS92 radiosonde systems [J].
Steinbrecht, Wolfgang ;
Claude, Hans ;
Schenenborn, Fritz ;
Leiterer, Ulrich ;
Dier, Horst ;
Lanzinger, Eckhard .
JOURNAL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY, 2008, 25 (06) :909-927
[4]   Systematic errors in global radiosonde precipitable water data from comparisons with ground-based GPS measurements [J].
Wang, Junhong ;
Zhang, Liangying .
JOURNAL OF CLIMATE, 2008, 21 (10) :2218-2238
[5]  
Performance of operational radiosondehumidity sensors in direct comparison with a chilledmirror dew-point hygrometer and its climate im-plication. Wang, J,D. J. Carlson,D. B. Parsons,T. F. Hock,D.Lauritsen,H. L. Cole,K. Beierle,E. Chamber-lain. Geophysical Research Letters . 2003