Quick Measurement of CH4, CO2 and N2O Emissions from a Short-Plant Ecosystem

被引:74
作者
王跃思
王迎红
机构
[1] Institute of Atmospheric Physics Chinese Academy of Sciences
[2] Beijing100029
关键词
gas chromatograph; valve; emission of greenhouse gases; ecosystem;
D O I
暂无
中图分类号
P402 [大气化学];
学科分类号
070602 ;
摘要
Combining improved injector, gas line and valve-driving models, a gas chromatograph (GC) equipped with Hydrogen Flame Ionization Detector (FID) and Electron Capture Detector (ECD), can measure CH4, CO2, and N2O simultaneously in an air sample in four minutes. Test results show that the system has high sensitivity, resolution, and precision; the linear response range of the system meets the requirement of flux measurements in situ. The system is suitable for monitoring fluxes of the main greenhouse gases in a short-plant field since it is easy to use, efficacious, and constant and reliable in collecting data.