Laparoscopic linear cutting stapler failure

被引:89
作者
Deng, DY
Meng, MV
Nguyen, HT
Bellman, GC
Stoller, ML
机构
[1] Univ Calif San Francisco, Sch Med, Dept Urol, San Francisco, CA 94143 USA
[2] Kaiser Permanente Med Ctr, Dept Urol, Los Angeles, CA 90034 USA
关键词
D O I
10.1016/S0090-4295(02)01778-8
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives. To characterize the frequency and nature of problems with linear cutting staplers to help prevent complications in the future. These devices are often used during laparoscopic urologic procedures. Methods. We retrospectively reviewed the experience with laparoscopic linear cutting staplers at two institutions routinely performing urologic laparoscopy and analyzed the difficulties with any staplers. Data from the Food and Drug Administration Center for Devices and Radiological Health were also examined to determine the prevalence and types of reported problems. Results. In performing approximately 460 laparoscopic cases, we encountered 5 problems (1%) with endovascular gastrointestinal anastomosis staplers. Fifty-five additional cases in 50 patients were documented in the Food and Drug Administration database. Of the 55 patients, 15 (27%) required open conversion to manage the problem, 8 (15%) received blood transfusions, and 2 (4%) died postoperatively. Twenty-two events occurred during 19 laparoscopic donor nephrectomies (35%) without associated graft dysfunction, damage, or loss. All phases of instrument use were subject to problems; however, abnormal firing of the stapler and improper staple formation were the most common and morbid aspects of device malfunction. Conclusions. Despite the general reliability of linear cutting staplers, difficulties were encountered in every step of use. Most situations were successfully managed by prompt identification and appropriate intracorporeal maneuvers. Nevertheless, significant morbidity may occur, and conversion to an open operation should be considered. Many potential problems can be avoided by surgeon and staff education, and one should be aware of the alternative methods of tissue ligation currently available.
引用
收藏
页码:415 / 419
页数:5
相关论文
共 9 条
[1]   Laparoscopic nephrectomy for renal cell cancer: Evaluation of efficacy and safety: A multicenter experience [J].
Cadeddu, JA ;
Ono, Y ;
Clayman, RV ;
Barrett, PH ;
Janetschek, G ;
Fentie, DD ;
McDougall, EM ;
Moore, RG ;
Kinukawa, T ;
Elbahnasy, AM ;
Nelson, JB ;
Kavoussi, LR .
UROLOGY, 1998, 52 (05) :773-777
[2]   Endovascular gastrointestinal stapler device malfunction during laparoscopic nephrectomy: Early recognition and management [J].
Chan, D ;
Bishoff, JT ;
Ratner, L ;
Kavoussi, LR ;
Jarrett, TW .
JOURNAL OF UROLOGY, 2000, 164 (02) :319-321
[3]   Laparoscopic radical nephrectomy: Cancer control for renal cell carcinoma [J].
Chan, DY ;
Cadeddu, JA ;
Jarrett, TW ;
Marshall, FF ;
Kavoussi, LR .
JOURNAL OF UROLOGY, 2001, 166 (06) :2095-2099
[4]  
Finlayson E V, 2001, Eff Clin Pract, V4, P172
[5]  
Jacobs SC, 2000, J UROLOGY, V164, P1494, DOI 10.1016/S0022-5347(05)67014-0
[6]  
Meng M V, 2001, Clin Transpl, P113
[7]   Improved recipient results after 5 years of performing laparoscopic donor nephrectomy [J].
Montgomery, RA ;
Kavoussi, LR ;
Su, LM ;
Sinkov, V ;
Cohen, C ;
Maley, WR ;
Burdick, JF ;
Markowitz, J ;
Ratner, LE .
TRANSPLANTATION PROCEEDINGS, 2001, 33 (1-2) :1108-1110
[8]   The long-term outcome of laparoscopic radical nephrectomy for small renal cell carcinoma [J].
Ono, Y ;
Kinukawa, T ;
Hattori, R ;
Gotoh, M ;
Kamihira, O ;
Ohshima, S .
JOURNAL OF UROLOGY, 2001, 165 (06) :1867-1870
[9]   Check the stapler before use [J].
Polglase, AL ;
Winnett, JD .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 1999, 69 (05) :402-403