Experience with a low profile bipolar, active fixation pacing lead in pediatric patients

被引:14
作者
Campbell, RM
Raviele, AA
Hulse, EJ
Auld, DO
McRae, GJ
Tam, VKH
Kanter, KR
机构
[1] Childresn Heart Ctr Egleston Scottish Rite, Atlanta, GA 30322 USA
[2] Emory Univ, Sch Med, Dept Cardiovasc Surg, Atlanta, GA USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1999年 / 22卷 / 08期
关键词
pacing; pediatrics; transvenous;
D O I
10.1111/j.1540-8159.1999.tb00594.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continued miniaturization of permanent pacing systems has promoted use of this technology in younger and smaller pediatric patients, Intermedics ThinLine 438-10 active fixation pacing leads (4.5 Fr lead body) were implanted in 26 patients (17 males/9 females; (9.9) over bar +/- 6.9 years). Twenty of 26 patients received dual chamber systems, 6 of 26 patients single lead systems. Each patient has been followed(3) 3 months. Pacemaker analysis at implant and 6 months later evaluated pulse width thresholds at 2.5 V (atrial <(0.07)over bar> +/- 0.02 vs <(0.13)over bar> +/- 0.02 ms [P = 0.01]; ventricular <(0.08)over bar> +/- 0.04 ms vs <(0.20)over bar> +/- 0.04 ms [P = 0.01]); sensing thresholds (atrial 4.1 +/- 0.41 mV vs 4.0 +/- 4.2 mV[P = NS]; ventricular (9.7) over bar +/- 0.72 vs (9.3) over bar +/- 0.94 mV [P = NS]); and impedance (atrial (345) over bar +/- 12 vs (370) over bar +/- 220 O [P = 0.04]; ventricular 412 +/- 17 vs 458 +/- 190 O [P < 0.01]). One volt lead failed with exit block at approximately 6 weeks. The youngest (9 months to 5 years) and smallest (6.5-18.0 kg) ten patients have each shown by venography to have at least mild venous stenosis at the lead(s) insertion site; five patients demonstrated collateral formation around asymptomatic obstruction, with no thrombus formation. The Intermedics 438-10 ThinLine pacing lead has demonstrated good and stable early postimplant electrical parameters. Angiographic evaluation in our smaller patients has shown evidence for asymptomatic venous obstruction.
引用
收藏
页码:1152 / 1157
页数:6
相关论文
共 18 条
[1]  
EPSTEIN ML, 1986, AM J CARDIOL, V57, P889, DOI 10.1016/0002-9149(86)90639-9
[2]   TRANSVENOUS PACEMAKERS IN CHILDREN - RELATION OF LEAD LENGTH TO ANTICIPATED GROWTH [J].
GHEISSARI, A ;
HORDOF, AJ ;
SPOTNITZ, HM .
ANNALS OF THORACIC SURGERY, 1991, 52 (01) :118-121
[3]   PEDIATRIC TRANSVENOUS PACING - A CONCERN FOR VENOUS THROMBOSIS [J].
GILLETTE, PC ;
ZEIGLER, V ;
BRADHAM, GB ;
KINSELLA, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1935-1939
[4]  
GILLETTE PC, 1995, PEDIAT CARDIAC PACIN
[5]   STEROID-ELUTING EPICARDIAL LEADS IN PEDIATRICS - IMPROVED EPICARDIAL THRESHOLDS IN THE 1ST YEAR [J].
HAMILTON, R ;
GOW, R ;
BAHORIC, B ;
GRIFFITHS, J ;
FREEDOM, R ;
WILLIAMS, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :2066-2072
[6]  
HAYES DL, 1983, J THORAC CARDIOV SUR, V85, P618
[7]   STEROID-ELUTING EPICARDIAL PACING LEADS IN PEDIATRIC-PATIENTS - ENCOURAGING EARLY RESULTS [J].
JOHNS, JA ;
FISH, FA ;
BURGER, JD ;
HAMMON, JW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (02) :395-401
[8]   Ten-year follow-up of a patient with pacemaker induced superior vena cava syndrome [J].
Kataoka, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (06) :1734-1736
[9]  
KUGLER J, 1989, AM HEART J, V3, P665
[10]   ACTUARIAL SURVIVAL OF TRANSVENOUS PACING LEADS IN A PEDIATRIC POPULATION [J].
LAU, YR ;
GILLETTE, PC ;
BUCKLES, DS ;
ZEIGLER, VL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (07) :1363-1367