In vitro laxity-testers for knee joints of mice

被引:14
作者
Blankevoort, L
vanOsch, GJVM
Janssen, B
Hekman, EEG
机构
[1] UNIV NIJMEGEN,INSTRUMENTS SERV DEV,6500 HB NIJMEGEN,NETHERLANDS
[2] UNIV NIJMEGEN,DEPT RHEUMATOL,6500 HB NIJMEGEN,NETHERLANDS
关键词
mouse; knee joint; laxity tester;
D O I
10.1016/0021-9290(95)00130-1
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The knee joints of mice can be used as a model for studying the effects of interventions on knee laxity. The goal of this study was to quantify knee joint laxity in vitro. Three devices were developed: a positioning- and cementing device, an anterior-posterior (AP) laxity tester and a varus-valgus (VV) laxity tester. The positioning- and cementing device was used to position the joint in a reproducible way and to attach clamping pins to the proximal femur and distal tibia using PMMA. The clamping pins were used to fix the joint to the AP- and VV-testers. In both testers the load was applied by means of a spindle-actuated spring while load and displacements were measured simultaneously. The load-displacement data were used to calculate displacement and compliance parameters. The performance of the testers was evaluated by testing 5 normal knee joints of 5 mice. Total AP-translation at +/-0.8 N was 0.43 (+/-0.16 S.D.) mm with compliances of 0.14 (+/-0.05 S.D.) mm N-1 and 0.12 (+/-0.05 S.D.) mmN(-1) at 0.8 N posterior and anterior force, respectively. Total VV-rotation at +/-4 Nmm was 17.2 (+/-2.6 S.D.) degrees with compliances of 0.9 degrees Nmm(-1) (+/-0.2 degrees Nmm(-1) S.D.) and 1.0 degrees Nmm(-1) (+/-0.4 degrees Nmm(-1) S.D.) at 4 Nmm valgus and varus moment, respectively. The contributions of the deformations of the bones and the fixtures to the rotations were negligible in the VV-test. In the AP-test they account for approximately 0.07 (+/-0.03 S.D.) mm of the total AP-translation. This will not affect the utilization of the device for comparative analysis. It is concluded that in vitro evaluation of AP- and VV-laxity in knees of mice is feasible with sufficient accuracy for evaluation of changes after ligament damage. (C) 1996 Elsevier Science Ltd.
引用
收藏
页码:799 / 806
页数:8
相关论文
共 16 条
[1]   ALLOGRAFT ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION IN A SHEEP MODEL - THE EFFECT OF SYNTHETIC AUGMENTATION [J].
AMENDOLA, A ;
FOWLER, P .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1992, 20 (03) :336-346
[2]  
[Anonymous], 1950, InNed. Akad. Wetensch. Proc. Ser. A
[3]   USE OF PATELLAR TENDON AUTOGRAFT FOR ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION IN THE RABBIT - A LONG-TERM HISTOLOGIC AND BIOMECHANICAL STUDY [J].
BALLOCK, RT ;
WOO, SLY ;
LYON, RM ;
HOLLIS, JM ;
AKESON, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1989, 7 (04) :474-485
[4]  
BRANDT KD, 1991, J RHEUMATOL, V18, P436
[5]  
DAWSONSAUNDERS B, 1990, BASIC CLIN BIOSTATIS
[6]   ACCURACY AND REPRODUCIBILITY OF INSTRUMENTED KNEE-DRAWER TESTS [J].
EDIXHOVEN, P ;
HUISKES, R ;
DEGRAAF, R ;
VANRENS, TJG ;
SLOOFF, TJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (03) :378-387
[7]   THE EPIDEMIOLOGY OF KNEE OSTEOARTHRITIS - RESULTS FROM THE FRAMINGHAM OSTEOARTHRITIS STUDY [J].
FELSON, DT .
SEMINARS IN ARTHRITIS AND RHEUMATISM, 1990, 20 (03) :42-50
[8]   OSTEOARTHROSIS IN A KNEE-JOINT DUE TO CHRONIC POSTTRAUMATIC INSUFFICIENCY OF THE MEDIAL COLLATERAL LIGAMENT - 9-YEAR FOLLOW-UP [J].
KANNUS, P ;
JARVINEN, M .
CLINICAL RHEUMATOLOGY, 1988, 7 (02) :200-207
[9]   POSTTRAUMATIC ANTERIOR CRUCIATE LIGAMENT INSUFFICIENCY AS A CAUSE OF OSTEO-ARTHRITIS IN A KNEE-JOINT [J].
KANNUS, P ;
JARVINEN, M .
CLINICAL RHEUMATOLOGY, 1989, 8 (02) :251-260
[10]   EXPERIMENTAL-MODEL OF OSTEOARTHRITIS - EARLY MORPHOLOGICAL AND BIOCHEMICAL CHANGES [J].
MCDEVITT, C ;
GILBERTSON, E ;
MUIR, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1977, 59 (01) :24-35