Correction factors for water-proofing sleeves in kilovoltage x-ray beams

被引:20
作者
Ma, CM [1 ]
Seuntjens, JP [1 ]
机构
[1] NATL RES COUNCIL CANADA,INST NATL MEASUREMENT STAND,IONIZING RADIAT STAND,OTTAWA,ON K1A 0R6,CANADA
关键词
radiotherapy; kilovoltage x-ray dosimetry; perturbation factor; waterproofing sleeve; EGS4 Monte Carlo;
D O I
10.1118/1.598039
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper investigates the effect of the waterproofing sleeve on the calibration of kilovoltage photon beams (50-300 kV). The sleeve effect correction factor, p(s), has been calculated using the Monte Carlo method as the ratios of the air kerma in an air cavity of a cylindrical chamber without the waterproofing sleeve to that with a sleeve. Three sleeve materials have been studied, PMMA, nylon and polystyrene. The calculations were carried out using the EGS4 (Electron Gamma Shower version 4) code system with the application of a correlated-sampling variance-reduction technique. The results show that the sleeve correction factor for a I-mm thick PMMA waterproofing sleeve surrounding a Farmer-type cylindrical chamber at 2 cm depth in water is about 0.991 at 50 kV (1.04 mm Al HVL) and 1.000 at 300 kV (21.5 mm Al HVL). For I-mm thick nylon and polystyrene sleeves, p(s) varies from 0.992 to 1.000 and from 0.981 to 1.000, respectively, for the same beam quality range. The p(s) factor varies with sleeve thickness, beam quality and phantom depth. No significant dependence of the p(s) factor on field size and source-surface distance has been found. Measurements for PMMA, nylon and polystyrene sleeves of various thicknesses have also been carried out and show excellent agreement with the Monte Carlo calculations. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1507 / 1513
页数:7
相关论文
共 20 条
[1]  
Andreo P., 1987, ABSORBED DOSE DETERM
[2]  
[Anonymous], P IAEA INT S MEAS AS
[3]   THE EFFECT OF THICKNESS OF THE WATERPROOFING SHEATH ON THE CALIBRATION OF PHOTON AND ELECTRON-BEAMS [J].
GILLIN, MT ;
KLINE, RW ;
NIROOMANDRAD, A ;
GRIMM, DF .
MEDICAL PHYSICS, 1985, 12 (02) :234-236
[4]   EFFECTS OF PLASTIC PROTECTIVE CAPS ON THE CALIBRATION OF THERAPY BEAMS IN WATER [J].
HANSON, WF ;
TINOCO, JAD .
MEDICAL PHYSICS, 1985, 12 (02) :243-248
[5]  
HOHLFELD K, 1996, IAEA TECHNICAL REPOR, V277, P77
[6]  
ICRU, 1984, 37 ICRU
[7]  
ICRU, 1973, 23 ICRU
[8]  
Klevenhagen SC, 1996, PHYS MED BIOL, V41, P2605, DOI 10.1088/0031-9155/41/12/002
[9]   CALCULATION OF ABSORBED DOSE RATIOS USING CORRELATED MONTE-CARLO SAMPLING [J].
MA, CM ;
NAHUM, AE .
MEDICAL PHYSICS, 1993, 20 (04) :1189-1199
[10]   MONTE-CARLO CALCULATED STEM EFFECT CORRECTIONS FOR NE2561 AND NE2571 CHAMBERS IN MEDIUM-ENERGY X-RAY-BEAMS [J].
MA, CM ;
NAHUM, AE .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (01) :63-72