A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement

被引:39
作者
Economos, CD
Nelson, ME
Fiatarone, MA
Dallal, GE
Heymsfield, SB
Wang, J
Yasumara, S
Ma, R
Vaswani, AN
RussellAulet, M
Pierson, RN
机构
[1] TUFTS UNIV,JEAN MAYER USDA,HUMAN NUTR RES CTR AGING,BOSTON,MA 02111
[2] ST LUKES ROOSEVELT HOSP,BODY COMPOSIT UNIT,NEW YORK,NY 10025
[3] BROOKHAVEN NATL LAB,UPTON,NY 11973
[4] WINTHROP UNIV HOSP,OSTEOPOROSIS RES CLIN,MINEOLA,NY 11501
[5] UNIV MICHIGAN,ANN ARBOR,MI 48109
关键词
dual energy X-ray absorptiometry; DXA; densitometry; body composition; lean mass; fat mass;
D O I
10.1038/sj.ejcn.1600400
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Objective: To assess intra- and inter-site soft tissue variability by dual energy X-ray absorptiometry (DXA). Design: Cross-sectional trial. Setting: Three medical research institutions. Subjects: Five humans (in vivo) and four phantoms (in vitro), configured from two whole body phantoms with artificial skeletons and thickness overlays. Interventions: Duplicate total-body DXA scans were performed on all subjects at each institution within a 15 d period. Results: All intra-site coefficients of variation (CV) were < 0.5% for total tissue mass, but in vitro and in vivo Cvs were 7.2% and 2.3% for fat mass (FM) and 2.5% and 0.9% for lean mass (LM), respectively. Several total-body and regional FM and LM measurements were significantly different between sites (P < 0.05), with percent differences between sites ranging from 2.6-13.3% for FM and from 1.6-13.6% for LM. Site 2 was consistently lower for FM and Site 3 was consistently lower for LM. Conclusions: These results stress the need for both rigorous and standardized cross-calibration procedures for soft tissue measurement by DXA.
引用
收藏
页码:312 / 317
页数:6
相关论文
共 47 条
[1]  
[Anonymous], SAS US GUID STAT
[2]   INTERSITE COMPARISON OF THE HOLOGIC QDR-1000 DUAL ENERGY X-RAY BONE DENSITOMETER [J].
BLAKE, GM ;
TONG, CM ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (761) :440-446
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   BODY-COMPOSITION AND BONE MASS IN POSTMENOPAUSAL WOMEN [J].
COMPSTON, JE ;
BHAMBHANI, M ;
LASKEY, MA ;
MURPHY, S ;
KHAW, KT .
CLINICAL ENDOCRINOLOGY, 1992, 37 (05) :426-431
[5]   REGIONAL CHANGES IN BODY-COMPOSITION BY TIME OF YEAR IN HEALTHY POSTMENOPAUSAL WOMEN [J].
DAWSONHUGHES, B ;
HARRIS, S .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1992, 56 (02) :307-313
[6]  
Economos CD, 1996, J BONE MINER RES, V11, P275
[7]   ACCURACY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY FOR BODY-COMPOSITION MEASUREMENTS IN CHILDREN [J].
ELLIS, KJ ;
SHYPAILO, RJ ;
PRATT, JA ;
POND, WG .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1994, 60 (05) :660-665
[8]   ASSESSMENT OF THE COMPOSITION OF MAJOR BODY REGIONS BY DUAL-ENERGY X-RAY ABSORPTIOMETRY (DEXA), WITH SPECIAL REFERENCE TO LIMB MUSCLE MASS [J].
FULLER, NJ ;
LASKEY, MA ;
ELIA, M .
CLINICAL PHYSIOLOGY, 1992, 12 (03) :253-266
[9]   4-COMPONENT MODEL FOR THE ASSESSMENT OF BODY-COMPOSITION IN HUMANS - COMPARISON WITH ALTERNATIVE METHODS, AND EVALUATION OF THE DENSITY AND HYDRATION OF FAT-FREE MASS [J].
FULLER, NJ ;
JEBB, SA ;
LASKEY, MA ;
COWARD, WA ;
ELIA, M .
CLINICAL SCIENCE, 1992, 82 (06) :687-693
[10]   DETECTION OF SMALL CHANGES IN BODY-COMPOSITION BY DUAL-ENERGY X-RAY ABSORPTIOMETRY [J].
GOING, SB ;
MASSETT, MP ;
HALL, MC ;
BARE, LA ;
ROOT, PA ;
WILLIAMS, DP ;
LOHMAN, TG .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1993, 57 (06) :845-850