Multicenter evaluation of the performance and clinical utility in longitudinal monitoring of the Bayer Immuno 1™ complexed PSA assay

被引:15
作者
Allard, WJ [1 ]
Cheli, CD
Morris, DL
Goldblatt, J
Pierre, Y
Kish, L
Chen, Y
Dai, J
Vessella, RL
Chan, DW
Schwartz, MK
Zhou, Z
Yeung, KK
机构
[1] Bayer Corp, Tarrytown, NY 10591 USA
[2] Univ Washington, Seattle, WA 98195 USA
[3] Johns Hopkins Univ, Baltimore, MD USA
[4] Mem Sloan Kettering Canc Ctr, New York, NY 10021 USA
关键词
prostate specific antigen; PSA-ACT; tumor markers; prostate cancer; performance; automation; immunoassays;
D O I
10.1177/172460089901400204
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
We conducted a multicenter evaluation of the analytical and clinical performance of the automated Bayer Immuno 1(TM) complexed PSA (cPSA) assay, and compared assay performance to the Bayer Immuno 1(TM) PSA assay. We sought to determine whether measurements of cPSA could be of clinical utility in the management of patients with prostate cancer. Results of the 10-day imprecision across three evaluation sites produced total CV < 2.50% and an analytical sensitivity of 0.02 mu g/L. There was an increased trend in clinical sensitivity for prostate cancer with increasing stage of disease (71-86%). Clinical specificity for patients with benign urogenital disease was 74.8%, and for other nonprostate diseases ranged from 91.1-100%. Retrospective serial monitoring of 155 patients with prostate cancer demonstrated concordance of cPSA measurements to clinical status for 97% of the patients analyzed. Results from the clinical studies using the Bayer Immuno 1 cPSA assay were comparable to results obtained with the Bayer Immuno 1 PSA assay. The Bayer Immuno 1 cPSA assay demonstrates analytical performance and clinical effectiveness in the management of prostate cancer patients during the course of disease and therapy.
引用
收藏
页码:73 / 83
页数:11
相关论文
共 45 条
  • [1] Allard WJ, 1998, CLIN CHEM, V44, P1216
  • [2] *AM JOINT COMM CAN, 1997, MAN STAG CANC, P219
  • [3] Measurement of complexed PSA improves specificity for early detection of prostate cancer
    Brawer, MK
    Meyer, GE
    Letran, JL
    Bankson, DD
    Morris, DL
    Yeung, KK
    Allard, WJ
    [J]. UROLOGY, 1998, 52 (03) : 372 - 378
  • [4] BRAWER MK, 1999, J UROL S, V161, P318
  • [5] SELECTION OF OPTIMAL PROSTATE-SPECIFIC ANTIGEN CUTOFFS FOR EARLY DETECTION OF PROSTATE-CANCER - RECEIVER OPERATING CHARACTERISTIC CURVES
    CATALONA, WJ
    HUDSON, MA
    SCARDINO, PT
    RICHIE, JP
    AHMANN, FR
    FLANIGAN, RC
    DEKERNION, JB
    RATLIFF, TL
    KAVOUSSI, LR
    DALKIN, BL
    WATERS, WB
    MACFARLANE, MT
    SOUTHWICK, PC
    [J]. JOURNAL OF UROLOGY, 1994, 152 (06) : 2037 - 2042
  • [6] Use of the percentage of free prostate-specific antigen to enhance differentiation of prostate cancer from benign prostatic disease - A prospective multicenter clinical trial
    Catalona, WJ
    Partin, AW
    Slawin, KM
    Brawer, MK
    Flanigan, RC
    Patel, A
    Richie, JP
    deKernion, JB
    Walsh, PC
    Scardino, PT
    Lange, PH
    Subong, ENP
    Parson, RE
    Gasior, GH
    Loveland, KG
    Southwick, PC
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1998, 279 (19): : 1542 - 1547
  • [7] CATALONA WJ, 1995, JAMA-J AM MED ASSOC, V274, P1214
  • [8] Chan Daniel W., 1996, Clinical Chemistry, V42, pS255
  • [9] Chan Daniel W., 1997, Journal of Urology, V157, P112
  • [10] Monoclonal antibodies 2F5 and 4G10 against prostate specific antigen (PSA) complexed to α1-antichymotrypsin
    Chen, ZX
    Chen, HB
    Stamey, TA
    [J]. JOURNAL OF UROLOGY, 1998, 160 (03) : 870 - 875