Transvascular imaging: Feasibility study using a vector phased array ultrasound catheter

被引:12
作者
Bruce, CJ [1 ]
Packer, DL [1 ]
Seward, JB [1 ]
机构
[1] Mayo Clin & Mayo Fdn, Div Cardiovasc Dis & Internal Med, Rochester, MN 55905 USA
来源
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES | 1999年 / 16卷 / 05期
关键词
transvascular; intravascular; ultrasound echocardiography; Doppler; interventional;
D O I
10.1111/j.1540-8175.1999.tb00086.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Transvascular imaging is defined as the acquisition of anatomic and functional information of structures lying beyond the confines of a vascular conduit within which the imaging device resides. Interrogating structures surrounding the vascular conduit is the subject of this feasibility study using a novel underblood, phased array ultrasound-tipped catheter. Methods: An intravascular catheter (10-F, 3.2-mm-diameter, four-way articulation) tipped with a 5.5- to 10-MHz frequency agile, vector phased array transducer with full Doppler capability (Sequoia, Acuson) was used. The imaging transducer has a wide range of tissue penetration (2 mm to >10 cm from the lens). The catheter was introduced via an 11-Fr femoral venous sheath into the inferior and superior vena cavae and right heart chambers. As the catheter was advanced attention, was directed to visualization of structures surrounding the vessel in which the catheter resided. Results: From the cavae and femoral vein, the thoracic, abdominal and femoral arteries could be easily imaged. Anatomy that was visualized included the liver, hepatic veins, gallbladder, and mesenteric vessels. Normal and pathological anatomy and Doppler physiology could be readily appreciated. Doppler (i.e., pulsed- and continuous-wave, color flow, and tissue Doppler) fostered unique transvascular physiological hemodynamic and flow assessment. Conclusion: Transvascular imaging is feasible in human subjects using this 10-Fr catheter tipped with a 5.5- to 10-MHz vector phased array transducer. Intravascular navigation to a desired Location within. the body and the performance of diagnostic or therapeutic procedures at a remote site under direct ultrasound visualization are possible. Full Doppler capability extends the concept of transvascular hemodynamic and physiological assessment.
引用
收藏
页码:425 / 430
页数:6
相关论文
共 27 条
[1]  
Beebe HG, 1998, J ENDOVASC SURG, V5, P101, DOI 10.1583/1074-6218(1998)005<0101:BCFDII>2.0.CO
[2]  
2
[3]   REAL-TIME ASSESSMENT OF EXPERIMENTAL ARTERIAL ANGIOPLASTY WITH TRANSVENOUS INTRAVASCULAR ULTRASOUND [J].
BULLER, CE ;
DAVIDSON, CJ ;
VIRMANI, R ;
KISSLO, K ;
SHEIKH, KH ;
STACK, RS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (01) :217-222
[4]   INTRACARDIAC ULTRASOUND MEASUREMENT OF VOLUMES AND EJECTION FRACTION IN NORMAL, INFARCTED, AND ANEURYSMAL LEFT-VENTRICLES USING A 10-MHZ ULTRASOUND CATHETER [J].
CHEN, CG ;
GUERRERO, JL ;
DEPRADA, JAV ;
PADIAL, LR ;
SCHWAMMENTHAL, E ;
CHEN, MH ;
JIANG, L ;
SVIZZERO, T ;
SIMON, H ;
THOMAS, JD ;
LEVINE, RA ;
WEYMAN, AE .
CIRCULATION, 1994, 90 (03) :1481-1491
[5]   RADIOFREQUENCY CATHETER ABLATION GUIDED BY INTRACARDIAC ECHOCARDIOGRAPHY [J].
CHU, E ;
FITZPATRICK, AP ;
CHIN, MC ;
SUDHIR, K ;
YOCK, PG ;
LESH, MD .
CIRCULATION, 1994, 89 (03) :1301-1305
[6]   Anatomy of atrioventricular nodal reentry investigated by intracardiac echocardiography [J].
DeLurgio, DB ;
Frohwein, SC ;
Walter, PF ;
Langberg, JJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (02) :231-234
[7]   Intracardiac echocardiography: In vitro and in vivo validation for right ventricular volume and function [J].
dePrada, JAV ;
Chen, MH ;
Guerrero, JL ;
Padial, LR ;
Jiang, L ;
Schwammenthal, E ;
Sagie, A ;
Weyman, AE ;
Levine, RA ;
Chen, CG .
AMERICAN HEART JOURNAL, 1996, 131 (02) :320-328
[8]   Determination of aortic valve area in valvular aortic stenosis by direct measurement using intracardiac echocardiography: A comparison with the gorlin and continuity equations [J].
Foster, GP ;
Weissman, NJ ;
Picard, MH ;
Fitzpatrick, PJ ;
Shubrooks, SJ ;
Zarich, SW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (02) :392-398
[9]   Usefulness of intracardiac echocardiography in complex transseptal catheterization during percutaneous transvenous mitral commissurotomy [J].
Hung, JS ;
Fu, M ;
Yeh, KH ;
Wu, CJ ;
Wong, P .
MAYO CLINIC PROCEEDINGS, 1996, 71 (02) :134-140
[10]   Quantitative assessment of stenotic aortic valve area by using intracardiac echocardiography: In vitro validation and initial in vivo illustration [J].
Jiang, L ;
DePrada, JAV ;
Lee, MY ;
Ba, JH ;
Padial, LR ;
Fallon, JT ;
King, ME ;
Palacios, IF ;
Weyman, AE ;
Levine, RA .
AMERICAN HEART JOURNAL, 1996, 132 (01) :137-144