Complications due to abandoned noninfected pacemaker leads

被引:57
作者
Böhm, A
Pintér, A
Duray, G
Lehoczky, D
Dudás, G
Tomcsányi, I
Préda, I
机构
[1] Semmelweis Univ, Fac Hlth Sci, Dept Med 2, H-1135 Budapest, Hungary
[2] Semmelweis Univ, Fac Hlth Sci, Dept Cardiovasc Surg, H-1135 Budapest, Hungary
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 12期
关键词
retained pacemaker lead; complication; noninfected; lead migration; skin erosion; muscle stimulation;
D O I
10.1046/j.1460-9592.2001.01721.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Noninfected unwanted pacemaker leads are usually abandoned since the reported complication rate related to them is low. We followed 60 patients with noninfected retained leads, and complication was observed in 12 (20%) of them, Lead migration occurred in 5 patients, skin erosion in 3 patients, venous thrombosis in 2 patients, and muscle stimulation in 2 patients. Management of the complications was a surgical procedure in seven patients, including two cases of open heart surgery, while chronic medical treatment was necessary in the other five patients, The results of this study suggest that complications due to noninfected abandoned leads may not be as rare as it was previously thought and may present a significant morbidity and cost burden. With the lead extraction technique available, the issue of the removal of all unwanted pacemaker leads should be addressed.
引用
收藏
页码:1721 / 1724
页数:4
相关论文
共 22 条
[1]   Muscle stimulation related to a nonfunctioning epicardial pacemaker electrode [J].
Bohm, A ;
Pintér, A ;
Hajdú, L ;
Préda, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (07) :1187-1188
[2]   Pacemaker lead fracture due to twiddler's syndrome [J].
Böhm, A ;
Komáromy, K ;
Pintér, A ;
Préda, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (05) :1162-1163
[3]   Cerebral embolism due to a retained pacemaker lead:: A case report [J].
Böhm, A ;
Bányai, F ;
Komáromy, K ;
Pintér, A ;
Préda, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (03) :629-630
[4]   The treatment of septicemia in pacemaker patients [J].
Bohm, A ;
Banyai, F ;
Preda, I ;
Zamolyi, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (07) :1105-1111
[5]   Peripheral pulmonary migration of a retained pacemaker lead [J].
Bohm, A ;
Komáromy, K ;
Pintér, A ;
Préda, I .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (08) :1272-1273
[6]  
BOHM A, 1998, J AM COLL CARDIOL SC, V31, P346
[7]  
Byrd CL, 1995, CLIN CARDIAC PACING, P491
[8]   AN UNUSUAL COMPLICATION OF THE UNREMOVED UNWANTED PACING WIRE [J].
DALAL, JJ ;
ROBINSON, CJ ;
HENDERSON, AH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (01) :14-16
[9]   UNUSUAL CASE OF PACEMAKER LEAD MIGRATION [J].
DALVI, BV ;
RAJANI, RM ;
LOKHANDWALA, YY ;
SATHE, SV ;
KULKARNI, HL ;
KALE, PA .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 21 (02) :95-96
[10]  
De Cock CC, 2000, PACE, V23, P423