A new method for rapid measurement of lactate in fetal and neonatal blood

被引:21
作者
Pennell, CE [1 ]
Tracy, MB [1 ]
机构
[1] Univ Sydney, Nepean Hosp, Dept Obstet & Gynaecol, Penrith, NSW 2751, Australia
关键词
D O I
10.1111/j.1479-828X.1999.tb03379.x
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
A prospective trial to determine the accuracy and precision of the Boehringer Mannheim Accusport(R) handheld lactate meter in measuring plasma lactate levels in umbilical cord blood and neonatal blood microsamples was performed in the labour ward and the neonatal intensive care unit of the Nepean Hospital. Specimens were collected from the umbilical artery of 160 consecutive deliveries covering gestations from 26 to 42 weeks, and from 110 umbilical artery catheters covering a range of gestations from 26 to 41 weeks. Serum was also obtained from an exchange transfusion for coefficient of variation analysis. Blood was simultaneously tested for lactic acid concentration on the Boehringer Mannheim (BM) Accusport(R) held lactate meter and the Radiometer ABL 625(R) blood-gas machine. Clinical data from the mother and baby were recorded together with the full blood-gas analysis for comparison with the lactate measures. Coefficients of variation for the BM Accusport(R) lactate meter were established by a further 120 samples of plasma lactate at 6 concentrations from 1 to 20 mmol/L. The stability of measurements with the BM lactate meter over a. wide range of temperatures was ascertained by repeated sampling of known concentrations of plasma lactate from 0.5 degrees C to 37 degrees C. The BM Accusport(R) lactate meter was found to be accurate from 1 mmol/L to 20 mmol/L with a Passing Bablok regression line y = 0.004 + 0.915 x (95% CI of slope of 0.889 to 0.946 and intercept -0.138 to 0.094) for whole blood, and y = 0.200 + 1.000 x (95% CI of slope 0.989 to 1.018 and intercept 0.080 to 0.222 for plasma. Between run coefficient of variation (CV) was calculated to be 1.23% to 5.53% over the clinically significant range (2.2-19.3 mmoVL). The BM lactate meter was accurate from 5 to 37 degrees C. At 0.5 degrees C the BM lactate meter significantly underestimated the plasma lactate concentration. There was no significant effect of haematocrit (41.5 to 62%), gestation, or operator on the accuracy of the BM lactate meter. The Accusport(R) handheld lactate meter is an accurate, commercially available, method of measuring plasma lactate levels in only 60 seconds at the point-of-care, It requires only 15 mu L of blood and is significantly cheaper than other assay methods. The BM lactate meter is well suited to assess lactic acidaemia of fetal scalp and neonatal blood samples to help quantify hypoxic stress in the perinatal period.
引用
收藏
页码:227 / 233
页数:7
相关论文
共 16 条
[1]   INTRAPARTUM FETAL SURVEILLANCE - THE ROLE OF FETAL SCALP BLOOD-SAMPLING [J].
CLARK, SL ;
PAUL, RH .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1985, 153 (07) :717-720
[2]   Association between blood lactate and acid-base status and mortality in ventilated babies [J].
Deshpande, SA ;
Platt, MPW .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1997, 76 (01) :F15-F20
[3]  
GAMBKE B, 1994, ACSM ANN M IND JUN 1
[4]  
GOODWIN TM, 1994, OBSTET GYNECOL, V83, P971
[5]   NEWBORN COMPLICATIONS AFTER INTRAPARTUM ASPHYXIA WITH METABOLIC-ACIDOSIS IN THE TERM FETUS [J].
LOW, JA ;
PANAGIOTOPOULOS, C ;
DERRICK, EJ .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1994, 170 (04) :1081-1087
[6]   THE CESAREAN-SECTION RATE IN SWEDEN - THE END OF THE RISE [J].
NIELSEN, TF ;
OLAUSSON, PO ;
INGEMARSSON, I .
BIRTH-ISSUES IN PERINATAL CARE, 1994, 21 (01) :34-38
[7]   LACTATE IN FETAL SCALP BLOOD AND UMBILICAL ARTERY BLOOD MEASURED DURING NORMAL LABOR WITH A TEST STRIP METHOD [J].
NORDSTROM, L ;
INGEMARSSON, I ;
PERSSON, B ;
SHIMOJO, N ;
WESTGREN, M .
ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 1994, 73 (03) :250-254
[8]   FETAL SCALP AND UMBILICAL ARTERY BLOOD LACTATE MEASURED WITH A NEW TEST STRIP METHOD [J].
NORDSTROM, L ;
PERSSON, B ;
SHIMOJO, N ;
WESTGREN, M .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1992, 99 (04) :307-309
[9]  
NORDSTROM L, 1995, THESIS STOCKHOLM U
[10]   CORRELATION OF NEONATAL ACID-BASE STATUS WITH APGAR SCORES AND FETAL HEART-RATE TRACINGS [J].
PAGE, FO ;
MARTIN, JN ;
PALMER, SM ;
MARTIN, RW ;
LUCAS, JA ;
MEEKS, GR ;
BUCOVAZ, ET ;
MORRISON, JC .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1986, 154 (06) :1306-1311