Experience with a urine opiate screening and confirmation cutoff of 2000 ng/mL

被引:13
作者
Fraser, AD
Worth, D
机构
[1] Queen Elizabeth II Hlth Sci Ctr, Toxicol Lab, Halifax, NS B3H 2Y9, Canada
[2] Dalhousie Univ, Halifax, NS, Canada
关键词
D O I
10.1093/jat/23.6.549
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Until recently, most laboratories used an opiate immunoassay screening and confirmation cutoff value of 300 ng/mL for codeine and morphine detection by gas chromatography-mass spectrometry (GC-MS). The cutoff value for opiates was increased to 2000 ng/mL or higher in various laboratories because of concerns that small doses of codeine and foods containing poppy seeds would give a positive opiate-screening result. Workplace drug-testing programs in the U.S. raised the opiate cutoff value to 2000 ng/mL on 30 November 1998. The objective of this study is to describe the results of opiate testing of 8600 urine specimens collected over 24 months with a 2000 ng/mL screening and confirmation (codeine and morphine) cutoff value. Specimens were screened by the EMIT®dau(TM) opiate assay using an in-house 2000-ng/mL morphine calibrator. Presumptive positive findings (N = 621) were analyzed quantitatively by GC-MS for codeine and morphine. One hundred and eighty six urine specimens were positive for codeine and morphine (> 2000 ng/mL), 298 specimens were positive for codeine only (> 2000 ng/mL) and 26 specimens were positive for morphine only (> 2000 ng/mL). All remaining specimens had codeine and morphine values < 2000 ng/mL. The codeine and morphine confirmation rate in this program reduced from 7.1% in 1994-1996 (300-ng/mL cutoff) to 2.1% in 1997-1998 with a 2000-ng/mL cutoff value. The codeine-only confirmation rate lowered from 6.6% (300-ng/mL cutoff) to 3.4% (2000-ng/mL cutoff). It was concluded that increasing opiate screening and codeine and morphine confirmation cutoff values led to > 300% reduction in the confirmed- positive rate for codeine and morphine and a 47% reduction in codeine-only confirmations in a urine drug-testing program where codeine was the major opiate used.
引用
收藏
页码:549 / 551
页数:3
相关论文
共 10 条
[1]  
*BECKM INSTR, CRE3 BECKM INSTR
[2]   GAS-CHROMATOGRAPHIC STUDY OF THE URINARY CODEINE-TO-MORPHINE RATIOS IN CONTROLLED CODEINE CONSUMPTION AND IN MASS-SCREENING FOR OPIATE DRUGS [J].
DUTT, MC ;
LO, DST ;
NG, DLK ;
WOO, SO .
JOURNAL OF CHROMATOGRAPHY, 1983, 267 (01) :117-124
[3]   POPPY SEED INGESTION AND OPIATES URINALYSIS - A CLOSER LOOK [J].
ELSOHLY, HN ;
ELSOHLY, MA ;
STANFORD, DF .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1990, 14 (05) :308-310
[4]  
ELSOHLY MA, 1995, HDB WORKPLACE DRUG T, P225
[5]  
Fraser AD, 1998, J FORENSIC SCI, V43, P194
[6]  
FRASER AD, 1998, CLIN BIOCHEM, V31, P299
[7]   CONFIRMATION OF MARIJUANA, COCAINE, MORPHINE, CODEINE, AMPHETAMINE, METHAMPHETAMINE, PHENCYCLIDINE BY GC MS IN URINE FOLLOWING IMMUNOASSAY SCREENING [J].
MULE, SJ ;
CASELLA, GA .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1988, 12 (02) :102-107
[8]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC STUDY OF CODEINE, NORCODEINE, AND MORPHINE AS INDICATORS OF CODEINE INGESTION [J].
POSEY, BL ;
KIMBLE, SN .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1984, 8 (02) :68-74
[9]  
1994, FED REG, V59, P29916
[10]  
1998, FED REG, V63, P63483