Calibration at different flow rates of a multistage liquid impinger

被引:38
作者
Asking, L
Olsson, B
机构
[1] PHARMACEUTICAL AND ANALYTICAL RESEARCH AND DEVELOPMENT, ASTRA DRACO AB, LUND, S-22100
关键词
D O I
10.1080/02786829708965456
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The multistage liquid impinger (MLI) is an impactor suitable for size-selective characterization and testing of pharmaceutical inhalation products, especially dry powder inhalers (DPIs). The four-stage version of the MLI, which was adopted by the European Pharmacopoeia for testing at 60 L/min, has been calibrated by two different laboratories (Astra Draco AB, Sweden, and AEA Technology, UK), reaching consistent results. The MLI has been extended with an additional impaction stage and all impaction stages have been calibrated at flow rates of 30, 40, 50, 60, 80, and 100 L/min. It is shown that in this flow range, the cutoff diameter depends on the how rate according to a simplified impaction equation, i.e., the cutoff diameter is proportional to the inverse of the square root of the flow rate. Based on the calibration data and the simplified equation, a cutoff value for any flow rate in the range 30-100 L/min can be assigned. This is an important possibility as a DPI should be tested at a how rate derived from the air flow resistance of the DPI. The cutoff values at 60 L/min for the four impaction stages are 13, 6.8, 3.1, and 1.7 mu m, the fifth stage being a filter stage of the five-stage MLI. The apparatus has also been shown to be robust and to have a high degree of uniformity between replicas. (C) 1997 American Association for Aerosol Research.
引用
收藏
页码:39 / 49
页数:11
相关论文
共 15 条
[1]  
[Anonymous], PHARMEUROPA
[2]   CALIBRATION AND USE OF THE AERODYNAMIC PARTICLE SIZER (APS 3300) [J].
BARON, PA .
AEROSOL SCIENCE AND TECHNOLOGY, 1986, 5 (01) :55-67
[3]  
BELL JH, 1973, J PHARM PHARMACOL, V25, pP32
[4]  
BYRON PR, 1994, PHARMACOPEIAL FORUM, V20, P7477
[5]   THE RELATIONSHIP BETWEEN POWDER INHALER RESISTANCE AND PEAK INSPIRATORY CONDITIONS IN HEALTHY-VOLUNTEERS - IMPLICATIONS FOR IN-VITRO TESTING [J].
CLARK, AR ;
HOLLINGWORTH, AM .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1993, 6 (02) :99-110
[6]  
GANDERTON D, 1996, PHARMEUROPA, V8, P245
[7]   A TECHNIQUE FOR ASSESSING SIZE-SELECTIVE DUST SAMPLERS USING THE APS AND POLYDISPERSE TEST AEROSOLS [J].
KENNY, LC ;
LIDEN, G .
JOURNAL OF AEROSOL SCIENCE, 1991, 22 (01) :91-100
[8]  
LODGE JP, 1986, CASCADE IMPACTOR
[9]  
MAISONNEUVE SA, 1995, EUROPEAN PHARMACOPEI, P8
[10]   MULTISTAGE LIQUID IMPINGER [J].
MAY, KR .
BACTERIOLOGICAL REVIEWS, 1966, 30 (03) :559-&