The self-calibration of a retarding-type Mott spin polarimeter with a large collection angle

被引:27
作者
Iori, K
Miyamoto, K
Narita, H
Sakamoto, K
Kimura, A [1 ]
Qiao, S
Shimada, K
Namatame, H
Taniguchi, M
机构
[1] Hiroshima Univ, Hiroshima Synchrotron Radiat Ctr, Hiroshima 7390046, Japan
[2] Hiroshima Univ, Grad Sch Sci, Hiroshima 7398526, Japan
关键词
D O I
10.1063/1.2162752
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
We have developed a compact retarding-type Mott spin polarimeter with a large collection angle at the Hiroshima Synchrotron Radiation Center and obtained a reliable value of the effective Sherman function by a self-calibration method. The spin polarization of secondary electrons was evaluated by extrapolation of the left-right scattering asymmetry at the zero energy-loss limit, which enabled us to obtain the effective Sherman function (S-eff) and the figure of merit of our polarimeter at different operating conditions. (c) 2006 American Institute of Physics.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 10 条
[1]
MOTT ELECTRON POLARIMETRY [J].
GAY, TJ ;
DUNNING, FB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (02) :1635-1651
[2]
EXTRAPOLATION PROCEDURES IN MOTT ELECTRON POLARIMETRY [J].
GAY, TJ ;
KHAKOO, MA ;
BRAND, JA ;
FURST, JE ;
MEYER, WV ;
WIJAYARATNA, WMKP ;
DUNNING, FB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :114-130
[3]
PRECISION-MEASUREMENT OF THE SHERMAN ASYMMETRY FUNCTION FOR ELECTRON-SCATTERING FROM GOLD [J].
GELLRICH, A ;
KESSLER, J .
PHYSICAL REVIEW A, 1991, 43 (01) :204-216
[4]
SIMPLE, COMPACT, MEDIUM-ENERGY MOTT POLARIZATION ANALYZER [J].
GRAY, LG ;
HART, MW ;
DUNNING, FB ;
WALTERS, GK .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (01) :88-91
[5]
NEW METHOD FOR ACCURATE CALIBRATION OF AN ELECTRON-SPIN POLARIMETER [J].
HOPSTER, H ;
ABRAHAM, DL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (01) :49-51
[6]
CALIBRATION OF A MOTT ELECTRON POLARIMETER - COMPARISON OF DIFFERENT METHODS [J].
MAYER, S ;
FISCHER, T ;
BLASCHKE, W ;
KESSLER, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (04) :952-957
[7]
USE OF THORIUM AS A TARGET IN ELECTRON-SPIN ANALYZERS [J].
MCCLELLAND, JJ ;
SCHEINFEIN, MR ;
PIERCE, DT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (04) :683-687
[8]
ABSOLUTE CALIBRATION OF A RETARDING-POTENTIAL MOTT POLARIMETER [J].
ORO, DM ;
BUTLER, WH ;
TANG, FC ;
WALTERS, GK ;
DUNNING, FB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) :667-669
[9]
A new compact electron spin polarimeter with a high efficiency [J].
Qiao, S ;
Kimura, A ;
Harasawa, A ;
Sawada, M ;
Chung, JG ;
Kakizaki, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (12) :4390-4395
[10]
CALIBRATION OF A MOTT DETECTOR USING CIRCULARLY POLARIZED IMPACT RADIATION FROM HELIUM [J].
UHRIG, M ;
BECK, A ;
GOEKE, J ;
ESCHEN, F ;
SOHN, M ;
HANNE, GF ;
JOST, K ;
KESSLER, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (05) :872-878