Outbreak of hemolytic reactions associated with chlorine and chloramine residuals in hemodialysis water

被引:11
作者
Calderaro, RVV
Heller, L
机构
[1] Secretaria Estado Saude Estado Minas Gerais, Belo Horizonte, MG, Brazil
[2] Univ Fed Minas Gerais, Dept Engn Sanitaria & Ambiental, Escola Engn, Belo Horizonte, MG, Brazil
来源
REVISTA DE SAUDE PUBLICA | 2001年 / 35卷 / 05期
关键词
hemodialysis; health surveillance; water analysis; chlorine; analysis; chloramine; water treatment; hemolytic reactions;
D O I
10.1590/S0034-89102001000500012
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Objective To investigate the process of water contamination and to assess the subsequent outbreak in the hemodialysis center Methods In September 2000, sixteen patients undergoing chronic hemodialysis tit a dialysis center in Minas Gerais, Brazil, experienced hemolytic reactions compatible with toxic symptoms due to chlorine and chloramine water contamination. Chlorine and chloramine concentrations in samples obtained from various sites of the dialysis center's water treatment and distribution system were measured. Case-patients were identified by reviewing medical records and nursing notes for all dialysis sessions carried out during the study period. Interviews with technicians, nursing and medical staff members were conducted. Results Reaction rate was significantly higher (p <0.028) during the outbreak period (September 25 to 27, 2000) than the pre-outbreak period (September 18 to 20, 2000). All patients with toxic symptoms had been under dialysis with water treated by reverse osmosis equipment and had used dialysers manually reprocessed. Chlorine and chloramine residuals concentrations found in the dialysis water as well as in the dialysers were tit levels higher than regulations, less than or equal to 0.5 mg/L for chlorine and less than or equal to 0.1 mg/L for chloramine. Individuals exposed to high chlorine and chloramine concentrations presented a relative risk of 2.58 (1.0-6.28) of having hemolytic reactions. Conclusion There is a need to observe surveillance procedures to secure that the maximum allowable concentrations in the water used in the hemodialysis process tire not exceeded.
引用
收藏
页码:481 / 486
页数:6
相关论文
共 11 条
[1]  
*AAMI, 1995, STAND REC PRACT, V3
[2]  
APHA, 1995, STAND METH WAT WAST
[3]  
BREGMAN H, 1996, MANUAL DIALISE, P139
[4]  
CALDERARO R, 1998, ENG SANIT AMB, V3, P118
[5]  
*COM EST NEFR SECR, 2000, REL MENS EST DIAL
[6]  
DEAN AG, 1995, EPI INFO VERSAO 6 CO
[7]   Chloramine-induced haemolysis presenting as erythropoietin resistance [J].
Fluck, S ;
McKane, W ;
Cairns, T ;
Fairchild, V ;
Lawrence, A ;
Lee, J ;
Murray, D ;
Polpitiye, M ;
Palmer, A ;
Taube, D .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (07) :1687-1691
[8]  
KAUFMAN AM, 1996, MANUAL DIALISE, P128
[9]   Erythropoietin resistance due to dialysate chloramine: the two-way traffic of solutes in haemodialysis [J].
Richardson, D ;
Bartlett, C ;
Goutcher, E ;
Jones, CH ;
Davison, AM ;
Will, EJ .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (11) :2625-2627
[10]   Quality of water used for haemodialysis:: bacteriological and chemical parameters [J].
Vorbeck-Meister, I ;
Sommer, R ;
Vorbeck, F ;
Hörl, WH .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (03) :666-675