Enhancement of in vitro effectiveness for hydrodynamic thrombectomy devices -: Simultaneous high-pressure rt-PA application

被引:9
作者
Müller-Hülsbeck, S [1 ]
Bathe, M [1 ]
Grimm, J [1 ]
Heller, M [1 ]
机构
[1] Univ Hosp Kiel, Dept Radiol, D-24105 Kiel, Germany
关键词
experimental interventional procedures; rheolytic/hydrodynamic thrombectomy; experimental thrombosis; mechanical thrombectomy; embolus weight; thrombolysis;
D O I
10.1097/00004424-199908000-00007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. TO determine the efficacy of simultaneous high-pressure recombinant tissue plasminogen activator (rt-PA) application for clot removal and procedure-related peripheral particle embolization for the hydrodynamic thrombectomy devices LF 140 AngioJet (LF 140) and triple-lumen Hydrolyser (triple HL) in an in vitro how model, METHODS. Thrombectomy of clots (n = 47) from 5-day-old human blood (8.9-9.7 g) was performed with the LF 140 and the triple HL, each with and without simultaneous rt-PA (9.97-12.59 g) application in a flow model (flow 1 liter/min) made of silicone tubes (7 mm inner tube diameter). All catheters were used according to the manufacturer's recommendations, RESULTS. The triple HL revealed no statistically significant performance differences with additional rt-PA, For the LF 140, mean thrombectomy time ranged from 23.5 sec (with rt-PA) to 33.5 sec (P = 0.05), The ratio of peripheral embolus weight to thrombus weight was reduced from 2.12% to 0.46% (rt-PA; P = 0.05). None of the tested devices had an isovolumetric performance; the mean ratio of applied saline to aspirated fluid for the devices were different from one, ranging from 0.89 to 0.92 (rt-PA) for the triple HL and from 0.43 to 0.52 (rt-PA) for the LF 140. No significant differences for remaining thrombus within the tubes were found, CONCLUSIONS. Simultaneous rt-PA-enhanced hydrodynamic thrombectomy is feasible in vitro. This combination reduces the time for thrombectomy and procedure-related peripheral particle embolization for the LF 140, No effect could be demonstrated for the triple HL in an in vitro how model. It remains unclear whether this procedure is effective in vivo, It seems likely that the incidence of fibrinolysis-associated complications may increase in vivo.
引用
收藏
页码:536 / 542
页数:7
相关论文
共 34 条
[1]   MECHANICAL CLOT DISSOLUTION - NEW CONCEPT [J].
BILDSOE, MC ;
MORADIAN, GP ;
HUNTER, DW ;
CASTANEDAZUNIGA, WR ;
AMPLATZ, K .
RADIOLOGY, 1989, 171 (01) :231-233
[2]   Comparative in vitro study of two percutaneous hydrodynamic thrombectomy systems [J].
Bucker, A ;
SchmitzRode, T ;
Vorwerk, D ;
Gunther, RW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (03) :445-449
[3]   MECHANICAL THROMBECTOMY - RESULTS OF EARLY EXPERIENCE [J].
COLEMAN, CC ;
KRENZEL, C ;
DIETZ, CA ;
NAZARIAN, GK ;
AMPLATZ, K .
RADIOLOGY, 1993, 189 (03) :803-805
[4]   RHEOLYTIC CATHETER FOR PERCUTANEOUS REMOVAL OF THROMBUS [J].
DRASLER, WJ ;
JENSON, ML ;
WILSON, GJ ;
THIELEN, JM ;
PROTONOTARIOS, EI ;
DUTCHER, RG ;
POSSIS, ZC .
RADIOLOGY, 1992, 182 (01) :263-267
[5]  
FOGARTY TJ, 1963, SURG GYNECOL OBSTET, V116, P241
[6]   RECANALIZATION OF OBSTRUCTED ARTERIES WITH A FLEXIBLE, ROTATING TIP CATHETER [J].
KENSEY, KR ;
NASH, JE ;
ABRAHAMS, C ;
ZARINS, CK .
RADIOLOGY, 1987, 165 (02) :387-389
[7]  
Kim RH, 1998, CATHETER CARDIO DIAG, V43, P483, DOI 10.1002/(SICI)1097-0304(199804)43:4<483::AID-CCD31>3.3.CO
[8]  
2-B
[9]   IMPROVED MECHANICAL THROMBOLYSIS WITH THE ADDITION OF RECOMBINANT TISSUE-TYPE PLASMINOGEN-ACTIVATOR - IN-VITRO STUDY WITH USE OF CR-51-LABELED CLOTS [J].
KURISU, Y ;
KAWAKAMI, K ;
TADA, S .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (06) :877-880
[10]  
LIEPSCH DW, 1986, BIORHEOLOGY, V23, P395