Lens distortion in optically coupled digital x-ray imaging

被引:8
作者
Liu, H [1 ]
Jiang, HY
Fajardo, LL
Karellas, A
Chen, WR
机构
[1] Johns Hopkins Univ, Dept Radiol & Biomed Engn, Baltimore, MD 21205 USA
[2] Univ Massachusetts, Med Ctr, Dept Radiol, Amherst, MA 01003 USA
[3] Univ Cent Oklahoma, Dept Phys & Engn, Edmond, OK 73034 USA
关键词
lens distortion; optically coupled x-ray imaging system; camera calibration;
D O I
10.1118/1.598956
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objectives of this research are to analyze geometrical distortions introduced by relay lenses in optically coupled digital x-ray imaging systems and to introduce an algorithm to correct such distortions. Methods: The radial and tangential errors introduced by a relay lens in digital x-ray imaging were experimentally measured, using a lens-coupled CCD (charge coupled device) prototype. An algorithm was introduced to correct these distortions. Based on an x-ray image of a standard calibration grid, the algorithm first identified the location of the optical axis, then corrected the radial and tangential distortions using polynomial transformation technique. Results: Lens distortions were classified and both radial and tangential distortions introduced by lenses were corrected using polynomial transformation. For the specific lens-CCD prototype investigated, the mean positional error caused by the relay lens was reduced by the correction algorithm from about eight pixels (0.69 mm) to less than 1.8 pixels (0.15 mm). Our investigation also shows that the fourth order of polynomial for the correction algorithm provided the best correction result. Conclusions: Lens distortions should be considered in position-dependant, quantitative x-ray imaging and such distortions can be minimized in CCD x-ray imaging by appropriate algorithm, as demonstrated in this paper. (C) 2000 American Association of Physicists in Medicine.
引用
收藏
页码:906 / 912
页数:7
相关论文
共 27 条
[1]  
ACHUELER B, 1996, SPIE P, V2432, P272
[2]  
[Anonymous], 1980, MANUAL PHOTOGRAMMETR
[3]   A three-step camera calibration method [J].
Bacakoglu, H ;
Kamel, MS .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (05) :1165-1172
[4]  
Ballard D.H., 1982, Computer Vision
[5]   ANALYSIS AND CORRECTION OF IMPERFECTIONS IN THE IMAGE INTENSIFIER-TV-DIGITIZER IMAGING CHAIN [J].
BOONE, JM ;
SEIBERT, JA ;
BARRETT, WA ;
BLOOD, EA .
MEDICAL PHYSICS, 1991, 18 (02) :236-242
[6]  
Brown D. C., 1966, PHOTOGRAMMETRIC ENG, V32, P444, DOI DOI 10.1234/12345678
[7]  
BROWN DC, 1971, PHOTOGRAMM ENG, V37, P855
[8]  
CASTLEMAN KR, 1996, DIGITAL IMAGE PROCES, P120
[9]   Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Correction of image intensifier distortion [J].
Fahrig, R ;
Moreau, M ;
Holdsworth, DW .
MEDICAL PHYSICS, 1997, 24 (07) :1097-1106
[10]   The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain [J].
Gronenschild, E .
MEDICAL PHYSICS, 1997, 24 (12) :1875-1888