Platelet Dysfunction in Outpatients With Left Ventricular Assist Devices

被引:110
作者
Steinlechner, Barbara [1 ]
Dworschak, Martin
Birkenberg, Beatrice
Duris, Monika
Zeidler, Petra
Fischer, Henrik
Milosevic, Ljubisa
Wieselthaler, Georg
Wolner, Ernst
Quehenberger, Peter
Jilma, Bernd
机构
[1] Med Univ Vienna, Div Cardiothorac & Vasc Anesthesia & Intens Care, Div Cardiothorac Surg, Inst Med, A-1090 Vienna, Austria
关键词
VON-WILLEBRAND-FACTOR; COAGULATION; AGGREGATION; EXPERIENCES; ACTIVATION; PFA-100; SUPPORT; CARE;
D O I
10.1016/j.athoracsur.2008.10.027
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Thromboembolic and bleeding complications in outpatients with a left ventricular assist device are common and can be detrimental. A meticulous balance between anticoagulant and procoagulant factors is therefore crucial. However, in contrast to routinely performed plasmatic coagulation tests, platelet function is hardly ever monitored although recent reports indicated platelet dysfunction. We therefore differentially evaluated platelet function with four commonly used point-of-care devices. Methods. In a cross-sectional design platelet function was assessed in 12 outpatients and 12 healthy matched volunteers using thrombelastography platelet mapping, thromboelastometry, platelet function analyzer, and a new whole blood aggregometer (Multiplate). Results. Phenprocoumon produced an international normalized ratio of 3.5. It was associated with a twofold prolongation in the thromboelastometry clotting time (p < 0.001). Platelet function under high shear was severely compromised: collagen adenosine diphosphate closure times were 2.5-fold longer in patients than in volunteers (p < 0.001), and 50% of patients had maximal collagen adenosine diphosphate closure time values. Although antigen levels of von Willebrand factor were 80% higher in patients (p < 0.001), von Willebrand factor-ristocetin was subnormal in 5 of 12 patients. Ristocetin-induced aggregation was also threefold higher in volunteers (p < 0.001), indicating an additional functional defect of platelets affecting the glycoprotein Ib-von Willebrand factor axis. The von Willebrand factor multimer pattern in patients also appeared abnormal. Conclusions. Multimodal antiplatelet monitoring showed markedly impaired platelet function in patients with a left ventricular assist device. Platelet dysfunction under high shear rates and abnormal ristocetin-induced aggregation is only partly attributable to low von Willebrand factor activity. These findings resemble the acquired von Willebrand syndrome that is associated with microaggregate formation and enhanced bleeding.
引用
收藏
页码:131 / 137
页数:7
相关论文
共 26 条
[1]   Plasma von Willebrand factor, thrombosis, and the endothelium: The first 30 years [J].
Blann, AD .
THROMBOSIS AND HAEMOSTASIS, 2006, 95 (01) :49-55
[2]  
CALATZIS A, 2006, BLOOD, V108
[3]   A novel modification of the thrombelastograph assay, isolating platelet function, correlates with optical platelet aggregation [J].
Craft, RM ;
Chavez, JJ ;
Bresse, SJ ;
Wortham, DC ;
Cohen, E ;
Carroll, RC .
JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 2004, 143 (05) :301-309
[4]  
DALE J, 1977, AM HEART J, V94, P359, DOI 10.1016/S0002-8703(77)80481-X
[5]   Analysis of platelet function during left ventricular support with the Incor and Excor system [J].
Etz, C ;
Welp, H ;
Rothenburger, M ;
Tjan, TDT ;
Wenzelburger, F ;
Schmidt, C ;
Scheld, HH ;
Schmid, C .
HEART SURGERY FORUM, 2004, 7 (05) :E423-E427
[6]   Coagulation monitoring and management of anticoagulation during cardiac assist device support [J].
Fries, D ;
Innerhofer, P ;
Streif, W ;
Schobersberger, W ;
Margreiter, J ;
Antretter, H ;
Hörmann, C .
ANNALS OF THORACIC SURGERY, 2003, 76 (05) :1593-1597
[7]   Coagulation monitoring: Current techniques and clinical use of viscoelastic point-of-care coagulation devices [J].
Ganter, Michael T. ;
Hofer, Christoph K. .
ANESTHESIA AND ANALGESIA, 2008, 106 (05) :1366-1375
[8]   Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease [J].
Geisen, Ulrich ;
Heilmann, Claudia ;
Beyersdorf, Friedhelm ;
Benk, Christoph ;
Berchtold-Herz, Michael ;
Schlensak, Christian ;
Budde, Ulrich ;
Zieger, Barbara .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2008, 33 (04) :679-684
[9]   CHARACTERIZATION OF THE UNIQUE MECHANISM MEDIATING THE SHEAR-DEPENDENT BINDING OF SOLUBLE VON-WILLEBRAND-FACTOR TO PLATELETS [J].
GOTO, S ;
SALOMON, DR ;
IKEDA, Y ;
RUGGERI, ZM .
JOURNAL OF BIOLOGICAL CHEMISTRY, 1995, 270 (40) :23352-23361
[10]   First experiences with a novel magnetically suspended axial flow left ventricular assist device [J].
Hetzer, R ;
Weng, YG ;
Potapov, EV ;
Pasic, M ;
Drews, T ;
Jurmann, M ;
Hennig, E ;
Müller, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 25 (06) :964-970