In vivo 1H magnetic resonance imaging and spectroscopy of the rat spinal cord using an inductively-coupled chronically implanted RF coil

被引:40
作者
Silver, X
Ni, WX
Mercer, EV
Beck, BL
Bossart, EL
Inglis, B
Mareci, TH
机构
[1] Univ Florida, Dept Biochem & Mol Biol, Gainesville, FL 32610 USA
[2] Univ Florida, Dept Phys, Gainesville, FL 32610 USA
[3] Univ Florida, Dept Neurosci, Gainesville, FL 32610 USA
[4] Univ Florida, McKnight Brain Inst, Gainesville, FL 32610 USA
关键词
implanted coils; MRI; NMR; spinal cord; spectroscopy;
D O I
10.1002/mrm.1319
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An inductively coupled, chronically implanted short-solenoid coil was used to obtain in vivo localized H-1 NMR spectra and diffusion-weighted images from a rat spinal cord. A 5 x 8 mm two-turn elliptically shaped solenoid coil was implanted in rats at the site of a T-12 vertebral-level laminectomy. Excitation was achieved solely by a 3 x 3 cm external surface coil, and signal detection was achieved by inductively coupling the external coil to the implanted coil. The image signal-to-noise ratio (SNR) obtained with the inductively-coupled implanted coil was compared with that obtained using a linear or a quadrature external surface coil. The implanted coil provided a gain by over a factor of 3 in SNR. The implanted coil was used to measure localized 1H spectra in vivo at the T13/L1 spinal-cord level within a 1.85 x 1.85 x 4.82 mm (16.5 muL) volume. With 256 averages, a similar to3-s repetition delay and respiratory gating, a high-quality spectrum was acquired in 13 min. In addition, water translational diffusion was measured in three orthogonal directions using a stimulated-echo imaging sequence, with a short echo time (TE), to produce a quantitative map of diffusion in a rat spinal cord in vivo. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:1216 / 1222
页数:7
相关论文
共 38 条
[1]   MAPPING OF METABOLITES IN WHOLE ANIMALS BY P-31 NMR USING SURFACE COILS [J].
ACKERMAN, JJH ;
GROVE, TH ;
WONG, GG ;
GADIAN, DG ;
RADDA, GK .
NATURE, 1980, 283 (5743) :167-170
[2]   Phosphorus-31 magnetic resonance spectroscopy studies of pig spinal cord injury - Myelin changes, intracellular pH, and bioenergetics [J].
Akino, M ;
ODonnell, M ;
Robitaille, PML ;
Stokes, BT .
INVESTIGATIVE RADIOLOGY, 1997, 32 (07) :382-388
[3]  
ANDERSON DK, 1991, RESTOR NEUROL NEUROS, V2, P309, DOI 10.3233/RNN-1991-245621
[4]  
ARAKAWA M, 1986, SOC MAGN RESON MED, V1, P199
[5]   Signal-to-noise ratio comparison between surface coils and implanted coils [J].
Arnder, LL ;
Shattuck, MD ;
Black, RD .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :727-733
[6]  
BECK BL, 1997, P 5 ISMRM, P1526
[7]  
BOSSART EL, 1999, THESIS U FLORIDA
[8]   ELECTRONIC DECOUPLING OF SURFACE-COIL RECEIVERS FOR NMR IMAGING AND SPECTROSCOPY [J].
EDELSTEIN, WA ;
HARDY, CJ ;
MUELLER, OM .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (01) :156-161
[9]   IMPLANTED COIL MR MICROSCOPY OF RENAL PATHOLOGY [J].
FARMER, THR ;
JOHNSON, GA ;
COFER, GP ;
MARONPOT, RR ;
DIXON, D ;
HEDLUND, LW .
MAGNETIC RESONANCE IN MEDICINE, 1989, 10 (03) :310-323
[10]   MAXIMIZING CONTRAST TO NOISE WITH INDUCTIVELY COUPLED IMPLANTED COILS [J].
FARMER, THR ;
COFER, GP ;
JOHNSON, GA .
INVESTIGATIVE RADIOLOGY, 1990, 25 (05) :552-558