Relationship of design features of stemmed tibial knee prosthesis with stress shielding and end-of-stem pain

被引:46
作者
Completo, A. [1 ]
Talaia, P. [1 ]
Fonseca, F. [2 ]
Simoes, J. A. [1 ]
机构
[1] Univ Aveiro, Dept Engn Mecan, P-3810193 Aveiro, Portugal
[2] Univ Beira Interior, Fac Ciencias Saude, P-6201001 Covilha, Portugal
关键词
ARTHROPLASTY; BONE; INTERFACE; COMPONENTS; CONTACT; CEMENT;
D O I
10.1016/j.matdes.2008.06.071
中图分类号
T [工业技术];
学科分类号
120111 [工业工程];
摘要
Materials and design are important issues within general product development and even more in biomedical product. Knee prosthesis is a product that still lacks of adequate design solutions. Stress shielding and stress concentrations related to the end-of-stem pain are frequent clinical evidences that are intimately related to the design of the prosthesis and steins used in revision total knee arthroplasty, to the femoral and tibial components. Stress shielding and stress concentrations were assessed through finite element analyses of different tibial stem designs. Stress shielding was determined considering the relative minimal principal cancellous bone stresses between implanted and intact tibiae models. Scintigraphy and radiographs of painful implanted knees were assessed to correlate clinical findings with the finite element stresses. This study showed that stem design, geometry and material change the mechanical behavior of bone around tibial tray and stem in revision TKA. For the analyzed bone regions, underneath tibial tray and around the stem, the geometry of stem had a more pronounced bone effect comparatively to the stem material. The results of this study support that short stems produce a minor effect in bone relatively to long stein in terms of stress shielding and stress concentration at tip region. No significant stress shielding differences was observed between Co-Cr and titanium stems. Overall, all stems provoked high stress concentrations in bone at the tip of the stem. Scintigraphy of a painful implanted knee showed local osteoblastic activity and radiographs evidence bone ossification at the distal tip of the stem due the stress concentration. The results show that long stems induce very high distal stresses and are more prone to induce pain. A new concept of long stein with a distal polymeric (flexible material) tip reduces the magnitude of stresses at the end of the stem and therefore can theoretically contribute to reduce pain. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1391 / 1397
页数:7
相关论文
共 18 条
[1]
THE EFFECT OF A STEM ON THE TIBIAL COMPONENT OF KNEE ARTHROPLASTY - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY OF UNCEMENTED TIBIAL COMPONENTS IN THE FREEMAN-SAMUELSON KNEE ARTHROPLASTY [J].
ALBREKTSSON, BEJ ;
RYD, L ;
CARLSSON, LV ;
FREEMAN, MAR ;
HERBERTS, P ;
REGNER, L ;
SELVIK, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (02) :252-258
[2]
The effect of stem design on end-of-stem pain in revision total knee arthroplasty [J].
Barrack, RL ;
Stanley, T ;
Burt, M ;
Hopkins, S .
JOURNAL OF ARTHROPLASTY, 2004, 19 (07) :119-124
[3]
Barrack RL, 1999, CLIN ORTHOP RELAT R, P216
[4]
BOURNE RB, 1986, CLIN ORTHOP RELAT R, P95
[5]
Evaluation of painful total knee arthroplasty [J].
Dennis, DA .
JOURNAL OF ARTHROPLASTY, 2004, 19 (04) :35-40
[6]
DYE SF, 1994, AAOS INSTR COURS LEC, V43, P453
[7]
REVISION TOTAL KNEE ARTHROPLASTY WITH USE OF MODULAR COMPONENTS WITH STEMS INSERTED WITHOUT CEMENT [J].
HAAS, SB ;
INSALL, JN ;
MONTGOMERY, W ;
WINDSOR, RE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (11) :1700-1707
[8]
KNEE ARTHROPLASTY USING A CEMENTLESS PCA PROSTHESIS WITH A POROUS-COATED CENTRAL TIBIAL STEM - CLINICAL AND RADIOGRAPHIC REVIEW AT 5 YEARS [J].
KIM, YH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (03) :412-417
[9]
KLEEMANN R, 2002, P 13 C EUR SOC BIOM, P223
[10]
Changes in bone density after cemented total knee arthroplasty - Influence of stem design [J].
Lonner, JH ;
Klotz, M ;
Levitz, C ;
Lotke, PA .
JOURNAL OF ARTHROPLASTY, 2001, 16 (01) :107-111