Validation of a high performance liquid chromatography analysis for the determination of noradrenaline and adrenaline in human urine with an on-line sample purification

被引:54
作者
Hansen, AM [1 ]
Kristiansen, J [1 ]
Nielsen, JL [1 ]
Byrialsen, K [1 ]
Christensen, JM [1 ]
机构
[1] Natl Inst Occupat Hlth, Dept Chem & Biochem, DK-2100 Copenhagen, Denmark
关键词
catecholamines; HPLC; on-line sample purification; human urine;
D O I
10.1016/S0039-9140(99)00029-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A high performance liquid chromatography (HPLC) method with fluorescence detection including an on-line purification was established for determination of catecholamines in human urine. The method was evaluated using samples of pooled urine spiked with catecholamines and validated for measurements of catecholamines in urine of healthy individuals in a held study. The laboratory method evaluation study showed that the recovery of the method was 0.82 (confidence interval (CI): 0.79-0.86) and 0.92 (CI: 0.89-0.95) for noradrenaline and adrenaline, respectively. Thus, correction factors of 0.82(-1) and 0.92(-1) were applied to correct the measurement results for this systematic effect. Furthermore, an uncertainty budget was generated for the analytical method using the BIPM-approach recommended by the International Organization for Standardization. The relative uncertainty of the method was estimated to be 10-12%, which was consistent with the observed relative variability found in the method evaluation. The method was evaluated in accordance with EURACHEM Guidance Document No 1 concerning accreditation for chemical laboratories with respect to accuracy and precision. The field study showed that the standard deviation of the method was sufficiently low to distinguish between persons working with two different processes of garbage collection, i.e. collection using four wheeled containers versus collection using bins. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:367 / 379
页数:13
相关论文
共 26 条
[1]  
BARTELS H, 1971, CLIN CHIM ACTA, V32, P81, DOI 10.1016/0009-8981(71)90467-0
[2]  
BAULIEU E, 1990, HORMONES MOL DIS, V1, P1
[3]   DEVELOPMENT AND PERFORMANCE OF AN AUTOMATED HPLC-ANALYZER FOR CATECHOLAMINES [J].
BOOS, KS ;
WILMERS, B ;
SAUERBREY, R ;
SCHLIMME, E .
CHROMATOGRAPHIA, 1987, 24 :363-370
[4]   EXTERNAL QUALITY ASSESSMENT OF URINARY PREGNANCY ESTROGEN ASSAY - FURTHER EXPERIENCE IN THE UNITED-KINGDOM [J].
BULLOCK, DG ;
WILDE, CE .
ANNALS OF CLINICAL BIOCHEMISTRY, 1985, 22 (MAY) :273-282
[5]   CAN GLYCOSYLATED HEMOGLOBIN BE A JOB STRESS PARAMETER [J].
CESANA, G ;
PANZA, G ;
FERRARIO, M ;
ZANETTINI, R ;
ARNOLDI, M ;
GRIECO, A .
JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 1985, 27 (05) :357-360
[6]   PROTOCOL FOR THE DESIGN AND INTERPRETATION OF METHOD EVALUATION IN ATOMIC-ABSORPTION SPECTROMETRIC ANALYSIS - APPLICATION TO THE DETERMINATION OF LEAD AND MANGANESE IN BLOOD [J].
CHRISTENSEN, JM ;
POULSEN, OM ;
ANGLOV, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (02) :329-334
[7]  
CHRISTENSEN JM, 1994, METHOD EVALUATION QU, V4, P45
[8]   APPLICATION OF A NEW AMIQAS COMPUTER-PROGRAM FOR INTEGRATED QUALITY-CONTROL, METHOD EVALUATION AND PROFICIENCY TESTING [J].
CHRISTENSEN, SL ;
ANGLOV, JTB ;
CHRISTENSEN, JM ;
OLSEN, E ;
POULSEN, OM .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 345 (2-4) :343-350
[9]  
*EURACHEM, 1993, EURACHEM GUID DOC 1, P1
[10]   A FULLY AUTOMATED CATECHOLAMINES ANALYZER BASED ON CARTRIDGE EXTRACTION AND HPLC SEPARATION [J].
GROSSI, G ;
BARGOSSI, A ;
LIPPI, A ;
BATTISTONI, R .
CHROMATOGRAPHIA, 1987, 24 :842-846