Evaluation of registration techniques for spinal image guidance

被引:39
作者
Holly, LT
Bloch, O
Johnson, JP
机构
[1] Cedars Sinai Med Ctr, Inst Spinal Disorders, Los Angeles, CA 90048 USA
[2] Univ Calif Los Angeles, Med Ctr, Dept Neurosurg, Los Angeles, CA 90024 USA
[3] Univ Calif San Francisco, Med Ctr, Dept Neurosurg, San Francisco, CA 94143 USA
关键词
image guidance; neuronavigation; registration error; pedicle screw; cadaver;
D O I
10.3171/spi.2006.4.4.323
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Paired point matching alone and paired point matching combined with surface matching are the two techniques used for the registration step in preoperative computerized tomography-based spinal image guidance. In the present study the authors sought to compare paired point-matching registration alone with paired point matching supplemented with surface matching to determine if the addition of surface matching improves navigational accuracy. Methods. Pedicle screws were placed in three embalmed human cervicothoracic spinal specimens during image guidance to serve as a reference points. The specimens were then rescanned, and each level was registered using paired point matching alone and then by paired point supplemented with Surface matching. Navigational accuracy was assessed by placing the stereotactic probe in the center of the screw head, and measuring the apparent distance between the screw head and probe on the computer monitor. Statistical analysis was used to compare the registration error and navigational error between the two techniques. Seventy-five screws were placed at 46 vertebral levels. The mean registration error for the paired point matching/surface matching technique (0.5 mm) was significantly lower (p < 0.001) than that of the paired point matching alone technique (1.2 mm); however, the intertechnique difference in navigational error was nearly equivalent (1.3 mm compared with 1.4 mm) and statistically insignificant (p > 0.05). Conclusions. Although the addition of surface matching to paired point registration significantly decreased the mean registration error, the actual navigational accuracy between the two techniques was equivalent when easily distinguishable points were meticulously selected. The use of paired point matching alone did not compromise the accuracy of navigation and is likely to result in decreased operating time.
引用
收藏
页码:323 / 328
页数:6
相关论文
共 24 条
[1]
Image-guided anterior cervical corpectomy - A feasibility study [J].
Albert, TJ ;
Klein, GR ;
Vaccaro, AR .
SPINE, 1999, 24 (08) :826-830
[2]
Effect of frameless stereotaxy on the accuracy of C1-2 transarticular screw placement [J].
Bloch, O ;
Holly, LT ;
Park, J ;
Obasi, C ;
Kim, K ;
Johnson, JP .
JOURNAL OF NEUROSURGERY, 2001, 95 (01) :74-79
[3]
EXTRACRANIAL APPLICATION OF THE FRAMELESS STEREOTAXIC OPERATING MICROSCOPE - EXPERIENCE WITH LUMBAR SPINE [J].
BRODWATER, BK ;
ROBERTS, DW ;
NAKAJIMA, T ;
FRIETS, EM ;
STROHBEHN, JW ;
KELLY, PJ ;
FRIEDMAN, WA .
NEUROSURGERY, 1993, 32 (02) :209-213
[4]
Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[5]
Foley KT, 1996, NEUROSURG CLIN N AM, V7, P171
[6]
Computer-aided pedicle screw placement using frameless stereotaxis [J].
Glossop, ND ;
Hu, RW ;
Randle, JA .
SPINE, 1996, 21 (17) :2026-2034
[7]
CLINICAL USE OF A FRAMELESS STEREOTAXIC ARM - RESULTS OF 325 CASES [J].
GOLFINOS, JG ;
FITZPATRICK, BC ;
SMITH, LR ;
SPETZLER, RF .
JOURNAL OF NEUROSURGERY, 1995, 83 (02) :197-205
[8]
Helm P A, 1998, Comput Aided Surg, V3, P51, DOI 10.1002/(SICI)1097-0150(1998)3:2<51::AID-IGS1>3.0.CO
[9]
2-J
[10]
Three-dimensional fluoroscopy-guided percutaneous thoracolumbar pedicle screw placement - Technical note [J].
Holly, LT ;
Foley, KT .
JOURNAL OF NEUROSURGERY, 2003, 99 (03) :324-329