Automatic measurement of vertebral rotation in idiopathic scoliosis

被引:28
作者
Adam, CJ
Askin, GN
机构
[1] Queensland Univ Technol, Paediat Spine Res Grp, Brisbane, Qld, Australia
[2] Mater Hlth Serv Brisbane Ltd, Brisbane, Qld, Australia
关键词
vertebral rotation; idiopathic scoliosis; automatic measurement; computed tomography; symmetry ratio algorithm;
D O I
10.1097/01.brs.0000197653.64796.9d
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Development of an automatic measurement algorithm and comparison with manual measurement methods. Objectives. To develop a new computer-based method for automatic measurement of vertebral rotation in idiopathic scoliosis from computed tomography images and to compare the automatic method with two manual measurement techniques. Summary of Background Data. Techniques have been developed for vertebral rotation measurement in idiopathic scoliosis using plain radiographs, computed tomography, or magnetic resonance images. All of these techniques require manual selection of landmark points and are therefore subject to interobserver and intraobserver error. Methods. We developed a new method for automatic measurement of vertebral rotation in idiopathic scoliosis using a symmetry ratio algorithm. The automatic method provided values comparable with Aaro and Ho's manual measurement methods for a set of 19 transverse computed tomography slices through apical vertebrae, and with Aaro's method for a set of 204 reformatted computed tomography images through vertebral endplates. Results. Confidence intervals (95%) for intraobserver and interobserver variability using manual methods were in the range 5.5 to 7.2. The mean (+/- SD) difference between automatic and manual rotation measurements for the 19 apical images was -0.5 degrees +/- 3.3 degrees for Aaro's method and 0.7 degrees +/- 3.4 degrees for Ho's method. The mean (+/- SD) difference between automatic and manual rotation measurements for the 204 endplate images was 0.25 degrees +/- 3.8 degrees. Conclusions. The symmetry ratio algorithm allows automatic measurement of vertebral rotation in idiopathic scoliosis without intraobserver or interobserver error due to landmark point selection.
引用
收藏
页码:E80 / E83
页数:4
相关论文
共 22 条
[1]
ESTIMATION OF VERTEBRAL ROTATION AND THE SPINAL AND RIB CAGE DEFORMITY IN SCOLIOSIS BY COMPUTER-TOMOGRAPHY [J].
AARO, S ;
DAHLBORN, M .
SPINE, 1981, 6 (05) :460-467
[2]
ROENTGENOGRAPHIC EVALUATION OF VERTEBRAL ROTATION [J].
BENSON, DR ;
SCHULTZ, AB ;
DEWALD, RL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (08) :1125-1129
[3]
Measurement of vertebral rotation in adolescent idiopathic scoliosis using three-dimensional magnetic resonance imaging [J].
Birchall, D ;
Hughes, DG ;
Hindle, J ;
Robinson, L ;
Williamson, JB .
SPINE, 1997, 22 (20) :2403-2407
[4]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]
BUNNELL WP, 1985, ORTHOP T, V9, P114
[6]
PRINCIPLES OF MEASUREMENT OF VERTEBRAL ROTATION FROM FRONTAL PROJECTIONS OF THE PEDICLES [J].
DRERUP, B .
JOURNAL OF BIOMECHANICS, 1984, 17 (12) :923-&
[7]
IMPROVEMENTS IN MEASURING VERTEBRAL ROTATION FROM THE PROJECTIONS OF THE PEDICLES [J].
DRERUP, B .
JOURNAL OF BIOMECHANICS, 1985, 18 (05) :369-378
[8]
Fait M, 1970, SCRIPTA MED, V43, P207
[9]
A new method to measure vertebral rotation from CT scans [J].
Göçen, S ;
Havitçioglu, H ;
Alici, E .
EUROPEAN SPINE JOURNAL, 1999, 8 (04) :261-265
[10]
Gocen S, 1998, J SPINAL DISORD, V11, P210