Temperature guided radiofrequency catheter ablation of myocardium: Comparison of catheter tip and tissue temperatures in vitro

被引:47
作者
Kongsgaard, E [1 ]
Steen, T [1 ]
Jensen, O [1 ]
Aass, H [1 ]
Amlie, JP [1 ]
机构
[1] NATL HOSP NORWAY,DEPT MED B,OSLO,NORWAY
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 05期
关键词
radiofrequency energy; catheter ablation; temperature control; temperature sensors;
D O I
10.1111/j.1540-8159.1997.tb06778.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Temperature monitoring during RF ablation has been proposed as a means of controlling the creation of the lesion. However, in vivo studies have shown poor correlation between lesion size and catheter Sip temperature. Thus, we hypothesized a difference between catheter tip and tissue temperatures during RF catheter ablation, and that this difference may depend on flow passing the ablation site, tip electrode length, and catheter-tissue orientation. In vitro studies were performed using four different ablation catheters (tip electrode length: 2, 4, or 6 mm) with a thermistor or a thermocouple as temperature sensor. Set temperature was 70 degrees C and pulse duration was 30 seconds. Pieces of porcine left ventricle were immersed in a bath of isotonic saline-dextrose solution at 37 degrees C. The ablation catheters were positioned perpendicularly, obliquely, or parallel to the endocardium. A temperature sensor was inserted from the epicardial side and positioned 1 mm beneath the catheter-tissue interface. Experiments were made with a flow of 200 mL/min passing the ablation site or with no flow. The catheter tip and tissue temperatures differed significantly (P < 0.0001) during ablation. This difference increased with time, with flow passing the ablation site, with the length of the tip electrode, and when the catheter was positioned perpendicularly or obliquely to the endocardium as compared to the parallel catheter-tissue orientation (P < 0.05). In conclusion, the tissue temperature may far exceed the catheter tip temperature, and intramyocardial superheating resulting in steam formation and popping may occur despite a relatively low catheter tip temperature.
引用
收藏
页码:1252 / 1260
页数:9
相关论文
共 12 条
[1]   ASSESSMENT OF EFFECTS OF A RADIOFREQUENCY ENERGY-FIELD AND THERMISTOR LOCATION IN AN ELECTRODE CATHETER ON THE ACCURACY OF TEMPERATURE-MEASUREMENT [J].
BLOUIN, LT ;
MARCUS, FI ;
LAMPE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (05) :807-813
[2]  
CARLSON LD, 1970, CONTROL ENERGY EXCHA, P55
[3]  
EASTOP TD, 1990, APPL THERMODYNAMICS, P671
[4]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[5]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[6]   RADIOFREQUENCY COAGULATION OF VENTRICULAR MYOCARDIUM - IMPROVED PREDICTION OF LESION SIZE BY MONITORING CATHETER TIP TEMPERATURE [J].
HINDRICKS, G ;
HAVERKAMP, W ;
GULKER, H ;
RISSEL, U ;
BUDDE, T ;
RICHTER, KD ;
BORGGREFE, M ;
BREITHARDT, G .
EUROPEAN HEART JOURNAL, 1989, 10 (11) :972-984
[7]   TEMPERATURE-CONTROLLED RADIOFREQUENCY CATHETER ABLATION OF AV CONDUCTION - 1ST CLINICAL-EXPERIENCE [J].
HOFFMANN, E ;
MATTKE, S ;
DORWARTH, U ;
MULLER, D ;
HABERL, R ;
STEINBECK, G .
EUROPEAN HEART JOURNAL, 1993, 14 (01) :57-64
[8]   POWER AND TEMPERATURE GUIDED RADIOFREQUENCY CATHETER ABLATION OF THE RIGHT ATRIUM IN PIGS [J].
KONGSGAARD, E ;
FOERSTER, A ;
AASS, H ;
MADSEN, S ;
AMLIE, JP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (10) :1610-1620
[9]   THE EFFECT OF TEMPERATURE-GUIDED RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
KONGSGAARD, E ;
FOERSTER, A ;
AASS, H ;
AMLIE, JP .
EUROPEAN HEART JOURNAL, 1993, 14 (06) :852-858
[10]   TEMPERATURE MONITORING DURING RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY PATHWAYS [J].
LANGBERG, JJ ;
CALKINS, H ;
ELATASSI, R ;
BORGANELLI, M ;
LEON, A ;
KALBFLEISCH, SJ ;
MORADY, F .
CIRCULATION, 1992, 86 (05) :1469-1474