Initial clinical experience with a video-based patient positioning system

被引:39
作者
Johnson, LS [1 ]
Milliken, BD [1 ]
Hadley, SW [1 ]
Pelizzari, CA [1 ]
Haraf, DJ [1 ]
Chen, GTY [1 ]
机构
[1] Univ Chicago, Dept Radiat & Cellular Oncol, Chicago, IL 60657 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 45卷 / 01期
关键词
patient positioning; immobilization; electronic portal imaging; head and neck; conformal radiotherapy;
D O I
10.1016/S0360-3016(99)00182-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report initial clinical experience with an interactive, video-based patient positioning system that is inexpensive, quick, accurate, and easy to use. Methods and Materials: System hardware includes two black-and-white CCD cameras, zoom lenses, and a PC equipped with a frame grabber. Custom software is used to acquire and archive video images, as well as to display real-time subtraction images revealing patient misalignment in multiple views. Two studies are described. In the first study, video is used to document the daily setup histories of 5 head and neck patients. Time-lapse cine loops are generated for each patient and used to diagnose and correct common setup errors. In the second study, 6 twice-daily (BID) head and neck patients are positioned according to the following protocol: at AM setups conventional treatment room lasers are used; at PM setups lasers are used initially and then video is used for 1-2 minutes to fine-tune the patient position. Lateral video images and lateral verification films are registered off-line to compare the distribution of setup errors per patient, with and without video assistance. Results: In the first study, video images were used to determine the accuracy of our conventional head and neck setup technique, i.e., alignment of lightcast marks and surface anatomy to treatment room lasers and the light field. For this initial cohort of patients, errors ranged from sigma = 5 to 7 mm and were patient-specific. Time-lapse cine loops of the images revealed sources of the error, and as a result, our localization techniques and immobilization device were modified to improve setup accuracy. After the improvements, conventional setup errors were reduced to sigma = 3 to 5 mm. In the second study, when a stereo pair of live subtraction images were introduced to perform daily "on-line" setup correction, errors were reduced to a = I to 3 mm. Results depended on patient health and cooperation and the length of time spent fine-tuning the position. Conclusion: An interactive, video-based patient positioning system was shown to reduce setup errors to within 1 to 3 mm in head and neck patients. without a significant increase in overall treatment time or labor-intensive procedures. Unlike retrospective portal image analysis, use of two live-video images provides the therapists with immediate feedback and allows for true 3-D positioning and correction of out-of plane rotation before radiation is delivered. With significant improvement in head and neck alignment and the elimination of setup errors greater than 3 to 5 mm, margins associated with treatment volumes potentially can be reduced, thereby decreasing normal tissue irradiation. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 18 条
[1]   ONLINE REPOSITIONING DURING TREATMENT OF THE PROSTATE - A STUDY OF POTENTIAL LIMITS AND GAINS [J].
BALTER, JM ;
CHEN, GTY ;
PELIZZARI, CA ;
KRISHNASAMY, S ;
RUBIN, S ;
VIJAYAKUMAR, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (01) :137-143
[2]   SETUP DEVIATIONS IN WEDGED PAIR IRRADIATION OF PAROTID-GLAND AND TONSILLAR TUMORS, MEASURED WITH AN ELECTRONIC PORTAL IMAGING DEVICE [J].
BEL, A ;
KEUS, R ;
VIJLBRIEF, RE ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (02) :153-159
[3]   PATIENT REPOSITIONING AND MOTION DETECTION USING A VIDEO CANCELLATION SYSTEM [J].
CONNOR, WG ;
BOONE, MLM ;
VEOMETT, R ;
HICKS, J ;
MILLER, RC ;
MAYER, E ;
SHEELEY, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1975, 1 (1-2) :147-153
[4]   ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH [J].
DENEVE, W ;
VANDENHEUVEL, F ;
DEBEUKELEER, M ;
COGHE, M ;
THON, L ;
DEROOVER, P ;
VANLANCKER, M ;
STORME, G .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) :45-54
[5]   INTERACTIVE USE OF ONLINE PORTAL IMAGING IN PELVIC RADIATION [J].
DENEVE, W ;
VANDENHEUVEL, F ;
COGHE, M ;
VERELLEN, D ;
DEBEUKELEER, M ;
ROELSTRAETE, A ;
DEROOVER, P ;
THON, L ;
STORME, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :517-524
[6]   WHEN AND HOW CAN WE IMPROVE PRECISION IN RADIOTHERAPY [J].
DUTREIX, A .
RADIOTHERAPY AND ONCOLOGY, 1984, 2 (04) :275-292
[7]   DAILY MONITORING AND CORRECTION OF RADIATION-FIELD PLACEMENT USING A VIDEO-BASED PORTAL IMAGING-SYSTEM - A PILOT-STUDY [J].
EZZ, A ;
MUNRO, P ;
PORTER, AT ;
BATTISTA, J ;
JAFFRAY, DA ;
FENSTER, A ;
OSBORNE, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (01) :159-165
[8]   A RANDOMIZED TRIAL OF PATIENT REPOSITIONING DURING RADIOTHERAPY USING A MEGAVOLTAGE IMAGING-SYSTEM [J].
GILDERSLEVE, J ;
DEARNALEY, DP ;
EVANS, PM ;
LAW, M ;
RAWLINGS, C ;
SWINDELL, W .
RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) :161-168
[9]   Automatic three-dimensional inspection of patient setup in radiation therapy using portal images, simulator images, and computed tomography data [J].
Gilhuijs, KGA ;
vandeVen, PJH ;
vanHerk, M .
MEDICAL PHYSICS, 1996, 23 (03) :389-399
[10]   IMMOBILIZATION ERROR - SOME THEORETICAL CONSIDERATIONS [J].
GOITEIN, M ;
BUSSE, J .
RADIOLOGY, 1975, 117 (02) :407-412